Using Gatan K2 Summit in Leginon » History » Version 2
Anchi Cheng, 05/29/2013 04:57 PM
1 | 1 | Anchi Cheng | h1. Using Gatan K2 Summit in Counting mode in Leginon |
---|---|---|---|
2 | |||
3 | If you want to use DM's dose fractionation, activate the "save frames" check box in the particular preset camera configuration. You can also use DM's frame alignment algorithm by activating "align frames" check box, but the latter is not recommended since the alignment in DM will delay the return of the image to Leginon and therefore affects throughput greatly. |
||
4 | |||
5 | 2 | Anchi Cheng | h2. Configuration |
6 | |||
7 | # Make sure your camera configuration is set to [[Leginon image orientation]] and instruments.cfg matches that resulting orientation and dimension. |
||
8 | # This is likely 270 degree rotated with a flip. New version of DM can acquire its internal darkware gain references in this configuration. |
||
9 | # K2's dose fractionation (frame saving) only works with full camera dimension without binning. |
||
10 | # K2 interface in DM restricts the exposure time to multiple of some set number, depending on the mode. When you enter in DM an invalid number, it automatically change it to the valid one when you click the acquire button. pyscope/dmsem.py has hard-coded values for these precision. Compare them to the precision of the DM version you have since it may change with DM version. |
||
11 | # K2 frame exposure time is adjustable in Leginon gui. However, you need to check if the number is acceptable by DM. |
||
12 | |||
13 | 1 | Anchi Cheng | h2. Preset Recommendation |
14 | |||
15 | See [[Pre-MSI_Set-up#Preset-Design-Example-for-small-pixel-Gatan K2 camera alone]] |
||
16 | |||
17 | h2. Calibrations |
||
18 | |||
19 | Most calibration and operation are similar to that of a typical digital camera. |
||
20 | |||
21 | Leginon treats linear/counted/super-resolution modes of K2 Summit as three cameras. Therefore, they each need its own calibrations. |
||
22 | A python script is available to make the copying of the calibrations easier. See message#1498. You may need to modify it for different modes you are copying calibrations to. |
||
23 | |||
24 | h2. Gain/Dark Correction Handling |
||
25 | |||
26 | 2 | Anchi Cheng | Leginon receives images from DM without software gain correction. This means gain correction is still needed in Leginon. |
27 | 1 | Anchi Cheng | |
28 | 2 | Anchi Cheng | h3. Counted/Super-Resolution |
29 | 1 | Anchi Cheng | |
30 | 2 | Anchi Cheng | # The hardware correction has been applied in DM before counting. Therefore, the "raw" image Leginon get are integers and roughly gain corrected. |
31 | # The dark image in Leginon should always be 0. Therefore, this need only be done once and no need to average several images in the making. |
||
32 | # The bright image acquired in Correction node should be taken at the intended exposure time such as 4-5 s, and a large number of images should be used to get an average image. This will account for variation in sensitivity of the pixels. There is likely a general gradient across the detector. |
||
33 | * movie frames recorded in dose-fractionation or frame-saving mode are "raw". Appion will use the bright/dark/correction plan of the image transferred to Leginon to make corrected frame during DD Frame Stack making. |
||
34 | 1 | Anchi Cheng | |
35 | 2 | Anchi Cheng | h3. Linear |
36 | 1 | Anchi Cheng | |
37 | 2 | Anchi Cheng | # The dark image will have a large mean and the value changes with exposure time. Ideally, the dark images should therefore be taken at the same exposure time as the later images to be corrected. Leginon can account for small differences but not a full range. |
38 | # To avoid saturation by the accumulated dark current, long exposure in linear mode is not recommended. 0.5 s is typical. |
||
39 | # Do not use frame saving with linear mode. It is not worth while. |
||
40 | 1 | Anchi Cheng | |
41 | 2 | Anchi Cheng | h2. Operation |
42 | 1 | Anchi Cheng | |
43 | 2 | Anchi Cheng | # Typical frame saving preset parameters: 0.2 s frame exposure time, 4-5 s total exposure time at 8-10 e/pixel/s (Some argue that 5 e/pixel/s is better). |
44 | # Because of its small size, we have a modified [[MSI_Quick-start_for_DD_only|Quick-start procedure]] we use regularly at NRAMM FEI F20. |
||
45 | # /myami/rawtransfer.py should be run on the file server to transfer the raw frames as it is produced. See [[DDD_raw_frame_file_transfer]]. K2 frames are saved as mrc stacks. |
||
46 | # See [[appion:Direct_Detector_Frame_Processing]] for tools used for processing the frames. |
||
47 | # To gain correct and align the frames, start parallel instances of Appion makeDDRawFrameStack.py See [[appion:GainDark correction of the raw frame with or without drift correction]]. |
||
48 | |||
49 | h2. Clean up |
||
50 | |||
51 | Frame saving generates a lot of files during acquisition and during processing. |
||
52 | |||