Actions
Using Gatan K2 Summit in Leginon » History » Revision 28
« Previous |
Revision 28/31
(diff)
| Next »
Anchi Cheng, 08/12/2014 10:33 AM
Using Gatan K2 Summit in Leginon¶
The main use of K2 Summit is in Counted or Super-resolution mode in combination with dose fractionation (frame saving).
If you want to use DM's dose fractionation, activate the "save frames" check box in the particular preset camera configuration. You can also use DM's frame alignment algorithm by activating "align frames" check box, but the latter is not recommended since the alignment in DM will delay the return of the image to Leginon and therefore affects throughput greatly.
- K2 Summit introduction
- Acquiring summed frame image thru Leginon
- Gain and Dark Correction This involves both DM and Leginon
- Pixel and Matrix Calibrations
- Trigger raw frame saving
- Transferring raw frame images off the hard drive of K2 Summit during data collection to a network drive.
- Compiling corrected movie stack with Appion using bright and dark images acquired in Leginon
Operation¶
- Typical frame saving preset parameters for counted/super-resolution: 0.2 s frame exposure time, 4-5 s total exposure time at 8-10 e/physical pixel/s (Some argues that 5 e/physical pixel/s is better). You can use DM's reading of dose rate while it is in counted mode and UNBINNed to get a reasonable estimate.
- Because of its small size, we have a modified Quick-start procedure we use regularly at NRAMM FEI F20.
- /myami/rawtransfer.py should be run on the file server to transfer the raw frames and give them names that are synced with the integrated image as it is produced. See DDD_raw_frame_file_transfer. K2 frames are saved as mrc stacks.
- See Direct_Detector_Frame_Processing for tools used for processing the frames.
- To gain correct and align the frames, start parallel instances of Appion makeDDRawFrameStack.py See GainDark correction of the raw frame with or without drift correction.
Warning¶
- Because of a delay in communication with DM when Leginon is started, camera configuration shown in Correction Node settings may not corresponds to what is used last time when you quit Leginon. Make sure you check this.
Leginon functions that do not work with K2 camera¶
General functions
- The dose matching tool in Preset Manager. K2's exposure time is not fine grained enough to make small adjustment according to the small dose deviation. You will have to use fixed exposure time and adjust beam intensity by hand to achieve the right dose. We find out the equivalent screen reading from the microscope for the acceptable 8-10 e/physical pixel/s dose rate for the detector and make adjustment there.
- Tomography correlation algorithm only takes square images. Until that is changed, Frame saving is not possible since only a full size unbinned image can be paired with frame saving.
Frame saving camera functions
- The image returned to Leginon display is always integrated over all the frames with K2 summit camera. Selection of frames to use as in DE series does not work here.
- Readout delay is not adjustable on K2. Leave it at 0.
Clean up¶
Frame saving generates a lot of files during acquisition and during processing.
- python script in leginon directory "cleanddraw.py" can help you clean up the transferred raw frames when you start to run out of disk space. See Feature #1784 for more details.
- Selective removal of aligned frame stack in Appion runs is still to be written.
Updated by Anchi Cheng about 10 years ago · 28 revisions