KLH Dataset II » History » Version 34
Amber Herold, 04/27/2012 01:43 PM
1 | 22 | Bridget Carragher | h1. Public Data Sets: KLH2 |
---|---|---|---|
2 | 21 | Bridget Carragher | |
3 | 23 | Eric Hou | *KLH Dataset II:* 655 defocus pairs of micrographs, coordinate centers of ~10000 side-view particles, a preliminary 3D reconstruction. |
4 | 1 | Eric Hou | |
5 | 23 | Eric Hou | h2. 1. Images. |
6 | 1 | Eric Hou | |
7 | 29 | Eric Hou | Images were acquired with a FEI Tecnai F20 equipped with a 2Kx2K CCD Tietz camera, as defocal pairs at a nominal magnification of 62,000 x and a voltage of 120 KeV, using the Leginon system ("Potter et al., 1999":http://www.ncbi.nlm.nih.gov/pubmed/10406132 ; "Carragher et al., 2000":http://www.ncbi.nlm.nih.gov/pubmed/11121305). The first image (named *.001.mrc) is acquired at near to focus (NTF) conditions (-0.6 to -1.5 µm) and the second image (named *.002.mrc) at farther from focus (FFF) conditions (-3µm). The time interval between the two exposures is approximately 20s due to the time required to read out the digital image from the camera. The pixel size is 2.2Å at the specimen scale and the accumulated dose for each high magnification image area was ~10 e/Ų. |
8 | 23 | Eric Hou | Below is an example of a defocus pair. Click on the images to see at full scale. |
9 | 2 | Eric Hou | |
10 | 4 | Eric Hou | p=. (a) !fig1a.jpg!:http://emg.nysbc.org/prtl_data/klh/klh_10k/fig1a.jpg (b) !fig1b.jpg!:http://emg.nysbc.org/prtl_data/klh/klh_10k/fig1b.jpg |
11 | 3 | Eric Hou | (a) Near to focus - NTF image. (b) Far from focus - FFF image. |
12 | |||
13 | 2 | Eric Hou | p=. *Figure 1: An example pair of high magnification images of KLH.* |
14 | |||
15 | 27 | Eric Hou | *Downloading High Magnification Images:* Image files are provided in "MRC":http://emg.nysbc.org/prtl_data/mrc_specification.htm format and as JPEG files |
16 | 19 | Eric Hou | |
17 | 34 | Amber Herold | Download entire set of images "here":http://emg.nysbc.org/prtl_data/klh/klh_10k/02oct09a.tar.gz (~83GB). |
18 | 6 | Eric Hou | |
19 | 25 | Eric Hou | We suggest using a tool called "em2em":http://www.imagescience.de/em2em, to convert MRC files to other formats. |
20 | |||
21 | 2 | Eric Hou | |
22 | |||
23 | 26 | Eric Hou | h2. 2. Positions of the Picked Particles in the Images |
24 | 2 | Eric Hou | |
25 | 26 | Eric Hou | Particles are picked in the FFF images and these coordinates are used to select particle in the NTF images. Phase correlation is used to align the two images to calculate the offset between the FFF coordinates and the NTF coordinates. |
26 | 2 | Eric Hou | |
27 | 10 | Eric Hou | p=. (a) !fig3a.jpg!:http://emg.nysbc.org/prtl_data/klh/klh_10k/fig3a.jpg (b) !fig3b.jpg!:http://emg.nysbc.org/prtl_data/klh/klh_10k/fig3b.jpg |
28 | 9 | Eric Hou | (a) The NTF image. (b) The FFF image. |
29 | |||
30 | 2 | Eric Hou | p=. *Figure 3: An example pair of images outlined with particles selected by Selexon. Each "+" indicate a detected particle.* |
31 | |||
32 | 26 | Eric Hou | h2. 2.1 Automated Picking |
33 | 12 | Eric Hou | |
34 | 30 | Eric Hou | The method described in ("Zhu et al., 2003":http://www.ncbi.nlm.nih.gov/pubmed/14643198) was used to select 11309 side-view particles from the FFF images. Particle coordinates are provide below: |
35 | 14 | Eric Hou | |
36 | 28 | Neil Voss | * "Coordinates":http://emg.nysbc.org/prtl_data/klh/klh_10k/selexon.pik.tar of 11309 side-view particles selected in the FFF images. |
37 | 15 | Eric Hou | |
38 | 1 | Eric Hou | |
39 | 26 | Eric Hou | h2. 3. 3D Reconstructed Map |
40 | 1 | Eric Hou | |
41 | 26 | Eric Hou | A 3D reconstruction was generated using automatically particles using D5 point-group symmetry and several cycles of refinement. |
42 | 1 | Eric Hou | |
43 | 28 | Neil Voss | * "Coordinates":http://emg.nysbc.org/prtl_data/klh/klh_10k/prtl_in_3dmap.tar of 10997 side-view particles went into the map. |