Actions
Bug #9472
openTIA and TIVIPS camera may not be reconnected through Leginon client when leginon application is switched without closing Leginon
Start date:
10/19/2020
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 3.5
Show in known bugs:
Yes
Workaround:
Restart Leginon client each time when Leginon applications need to be changed.
Description
TVIPS camera interface has this problem for a long time now. TIA driven camera (Ceta) starts having the same with 64-bit comtypes 1.1.7. The latter also gives segmentation error when closing dll connection.