Project

General

Profile

Actions

Support #1200

open

normalization

Added by Scott Stagg over 13 years ago. Updated over 13 years ago.

Status:
Assigned
Priority:
Normal
Assignee:
Category:
-
Target version:
-
Start date:
02/24/2011
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.1.0

Description

I am having trouble with my autofocus, and there always seems to be a systematic offset of around 1.5 microns more than the defocus I tell leginon to set. I think I have tracked it down to a normalization problem. I want to try adding in an objective lens normalization right before the autofocus. I see what function to call in tecnai.py in pyscope, but I don't see how to use it. Do you have a suggestion for the best way to do this?

Thanks,
Scott