Bug #12775
closed
The next application launched after autoscreen starts off and close column valve automatically if just kill and run the same application
Added by Anchi Cheng over 2 years ago.
Updated about 2 years ago.
Affected Version:
Appion/Leginon 3.6
- Subject changed from The next application launched after autoscreen starts off automatically if leginon was not reopened to The next application launched after autoscreen starts off and close column valve automatically if just kill and run the same application
Main problem is not resetting broadcast list when killing the node. PresetsManager also not having lock on idle_active if two events are settings it.
- Status changed from Assigned to Closed