Project

General

Profile

Actions

Bug #1445

closed

Bug #1048: GIF switches to EFTEM mode, has different magnifications

two retractable camera on the same scope causes constant inserting/retracting

Added by Anchi Cheng about 13 years ago. Updated about 7 years ago.

Status:
Won't Fix or Won't Do
Priority:
Low
Assignee:
Category:
-
Target version:
-
Start date:
10/11/2011
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.2.0
Show in known bugs:
No
Workaround:

Change what is returned in pyscope.gatan.Gatan.getRetractable() function to be False all the time.


Description

Reported by Scott Stagg. With their two gatan camera GatanGIF and TecnaiEFTEM and instrument.cfg that includes both, Leginon try to insert and retract them between acquisitions even though only one is used in the presets.

See workaround.

The bug appears since we make it to detect and switch camera for DD.

Actions

Also available in: Atom PDF