Project

General

Profile

Actions

Bug #1795

closed

manual focus tool in focuser should not use the first preset in preset order

Added by Anchi Cheng over 12 years ago. Updated over 12 years ago.

Status:
Closed
Priority:
Urgent
Assignee:
Category:
Leginon subsystem
Start date:
05/14/2012
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.2.0 (trunk)
Show in known bugs:
No
Workaround:

Description

This is a problem when manual focus should be done with survey camera on DD which does not have a valid matrix calibration. When image shift is used as move type, it uses first preset in preset order to determine the transformation of the target. The two usage conflict with each other.

Will create a new settings for manual focus. Need dbschema update for all users. and need to be in 2.2