Bug #2517

closed

Corrector instruments in settings gui does not match what Leginon think it is on when initialized

Added by Anchi Cheng about 11 years ago. Updated almost 11 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
Leginon GUI
Start date:
09/12/2013
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.2.0
Show in known bugs:
No
Workaround:

Description

Instrument in settings are typically shown as the one last loaded but GetValue of the widget gives what is last saved in the database. This is because Loading instrument often is delayed so that the widget was already set.

This causes a confusion since we will see the last loaded instrument selected but when we close the settings dialog and collect the reference image, they would be of the instrument in the database since the instrument are not set when the dialog is closed but are set when the selector choice is changed.

This problem was not obvious until we start having multiple camera on the same scope and different tem classes set for regular TEM vs EFTEM.


Related issues 1 (0 open1 closed)

Has duplicate Leginon - Bug #2516: Reference images saved as a different instrument than normal acquisitionClosedAnchi Cheng09/11/2013