Project

General

Profile

Actions

Bug #2561

closed

Tomography stage tilt not reset after previous tilt-series aborted

Added by Jian Shi almost 11 years ago. Updated over 6 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
10/18/2013
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.2.0
Show in known bugs:
No
Workaround:

Description

SVN Version17944. Application:tomography

The tilt-series can be aborted during the middle of data collection, mostly due to low intensity or block of beam. However the stage tilt angle will stay as where was aborted. Then tomography still carry on to the next position, but with that tilted angle instead of 0 for navigation. The targeting will not be accurate due to stage tilt. The tilt-series will be collected at wrong places.

I notice such issue long time ago, and thought I reported it before.

Actions

Also available in: Atom PDF