SVN Version17944. Application:tomography

The tilt-series can be aborted during the middle of data collection, mostly due to low intensity or block of beam. However the stage tilt angle will stay as where was aborted. Then tomography still carry on to the next position, but with that tilted angle instead of 0 for navigation. The targeting will not be accurate due to stage tilt. The tilt-series will be collected at wrong places.

I notice such issue long time ago, and thought I reported it before.