Actions
Bug #3157
closedautomated apertures can not be controlled from scripting at NYSBC 3200fsc
Status:
Won't Fix or Won't Do
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
04/24/2015
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 3.2
Show in known bugs:
Yes
Workaround:
set USE_AUTO_APT=False in jeol.cfg or access scripting from Jeol computer instead.
Description
Until we fix this, needs a mechanism to keep Leginon from giving errors.
Updated by Anchi Cheng over 9 years ago
- Is duplicate of Bug #3026: jeol automated aperture not always accessable through scripting added
Updated by Anchi Cheng about 9 years ago
- Subject changed from automated apertures can not be controlled from scripting on NYSBC 3200 to automated apertures can not be controlled from scripting at NYSBC 3200fsc
- Workaround updated (diff)
It is a problem of external access of TEM3 scripting. The problem went away once we control it from Jeol computer.
Updated by Anchi Cheng about 9 years ago
- Status changed from Assigned to Won't Fix or Won't Do
- Target version set to Appion/Leginon 3.2
- Affected Version changed from Appion/Leginon 3.1.0 to Appion/Leginon 3.2
- Show in known bugs changed from No to Yes
- Workaround updated (diff)
I can not fix this instrument problem. Use workaround
Actions