Description
GatanK2Super camera has only the full size camera configuration which is not suitable for calibration.
It is better to copy the calibrations from GatanCounting over.
http://emg.nysbc.org/boards/6/topics/1496?r=1498#message-1498 contains a version of this called copycal.py
r18859 gives this a way to use this without the need to modify the script source code.
1. Run a test first. Make sure it does not give errors.
python copy_k2_super_cal.py hostname high_tension
2. Add commit flag 1 when rerun to commit the change, for example,
python copy_k2_super_cal.py hostname 200000 1