Project

General

Profile

Actions

Bug #5059

closed

stage alpha position slip over time

Added by Anchi Cheng over 7 years ago. Updated about 7 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
07/01/2017
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 3.2
Show in known bugs:
Yes
Workaround:

Description

It has become more common on Krios that the stage wobble tilt does not truly return to the original angle after eucentric focusing by stage tilting. In some case each iteration causes 0.03 degree of change. Since such as small value is within the resolution of the control, it does not get corrected. However, this accumulates over many iterations and can become significant.


Related issues 1 (0 open1 closed)

Related to Leginon - Bug #5060: tilt like parent does not work if the tilt is modified by focusing, alignment, or conditioningClosedAnchi Cheng07/02/2017