Description

To save time for calibrations. This is especially useful for stage position calibration since such calibration is impossible at high magnifications and yet the dirctionality provided by the matrix is needed in operation.

See Scaling_Stage_Position_Matrix_Calibration

Actions #1

Updated by Anchi Cheng about 7 years ago

First round is a script for stage position matrix. Don't want to do all matrix type, yet.

See also Scaling_Stage_Position_Matrix_Calibration

Actions #2

Updated by Anchi Cheng about 7 years ago

  • Description updated (diff)
Actions #3

Updated by Anchi Cheng about 7 years ago

  • Status changed from Assigned to In Test
  • Target version changed from Appion/Leginon Future Version to Appion/Leginon 3.3

Replaced matrix calibrator pixel to pixel tool which is not used any more with this function. It does to all matrixes available there. This is a short cut without considering the possibility that image orientation may not be the same, but it is a good first assumption for stage position, any way.

Actions #4

Updated by Anchi Cheng almost 7 years ago

  • Status changed from In Test to Closed

Works.

If there is lens program change that changes image rotation, calibrate on the lowest mag of the new lens program and then scale from there.