Bug #5281
open
tomography tilt correlator shift correction assume tilt axis is y
Added by Anchi Cheng about 7 years ago.
Updated almost 7 years ago.
Affected Version:
Appion/Leginon 3.2
Description
An old quick change that was never fixed
- Status changed from Assigned to In Test
- Assignee changed from Anchi Cheng to Alex Noble
Alex, please try it out with the trunk version. If this correct, we should see improvement of Krios 1 when using the option "Stretch images according to the tilt before correlation" The way it was before would not change much since the shift in non-tilt axis direction is small anyway.
This still does not behaves right. tiltcorrelator.TiltCorrelator.tiltShift function is making scopes with ~90 deg tilt axis phi to affect more in x axis shift. Remove all tiltShift correction now. Will reinvestigate later.