Project

General

Profile

Actions

Bug #5387

open

Ceta error - unable to get corrected image

Added by Mykhailo Kopylov almost 7 years ago. Updated almost 7 years ago.

Status:
Assigned
Priority:
Normal
Assignee:
Category:
-
Start date:
12/27/2017
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 3.3
Show in known bugs:
No
Workaround:

Restart camera client and Leginon.


Description

During preset alignments, when taking gr image through navigator Ceta camera inserts properly but doesn't take an image.
The error in Leginon says "unable to get corrected image". Image doesn't appear in TIA or Leginon.
Presets that are not using Ceta work properly.

Fixed by restarting camera client and Leginon.

This happened for the first time in session 17dec26h. The issue was replicated in session 17dec27a during normal preset alignments.


Files

17dec27a_leginon_error.png (1.53 MB) 17dec27a_leginon_error.png Mykhailo Kopylov, 12/27/2017 11:51 AM
Actions #1

Updated by Anchi Cheng almost 7 years ago

Both of these are on Krios2. It may have to do with the software upgrade. I will need scope time to figure this out.

Actions #2

Updated by Anchi Cheng almost 7 years ago

  • Status changed from New to Assigned
  • Assignee changed from Mykhailo Kopylov to Anchi Cheng
  • Priority changed from Normal to Urgent
Actions #3

Updated by Anchi Cheng almost 7 years ago

  • Category deleted (Leginon GUI)

Misha and I try to reproduce but were unable to. Both incidence happened after Cs corrector tuning, so it may be related since that part uses Falcon but EF-TEM does not include Falcon.

Actions #4

Updated by Anchi Cheng almost 7 years ago

  • Priority changed from Urgent to Normal
Actions