Actions
Bug #5387
openCeta error - unable to get corrected image
Status:
Assigned
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
12/27/2017
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 3.3
Show in known bugs:
No
Workaround:
Restart camera client and Leginon.
Description
During preset alignments, when taking gr image through navigator Ceta camera inserts properly but doesn't take an image.
The error in Leginon says "unable to get corrected image". Image doesn't appear in TIA or Leginon.
Presets that are not using Ceta work properly.
Fixed by restarting camera client and Leginon.
This happened for the first time in session 17dec26h. The issue was replicated in session 17dec27a during normal preset alignments.
Files