Adding DE-12 to Leginon » History » Version 10
Anchi Cheng, 01/09/2014 09:32 PM
1 | 9 | Anchi Cheng | h1. Installation Guide for DE-12 or DE-20 |
---|---|---|---|
2 | 1 | Anchi Cheng | |
3 | 9 | Anchi Cheng | h2. DE-12 is on a computer separated from the microscope |
4 | 1 | Anchi Cheng | |
5 | 9 | Anchi Cheng | Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first. |
6 | 1 | Anchi Cheng | |
7 | 9 | Anchi Cheng | h2. Extra Package and Installation |
8 | |||
9 | None. |
||
10 | |||
11 | h2. instruments.cfg |
||
12 | |||
13 | {{include(Instrument.cfg for DE-12)}} |
||
14 | height and width should be set according to the model. |
||
15 | |||
16 | h2. Setup |
||
17 | |||
18 | * Create the folder to store the dose fractionation raw frames. Default location is D:\\frames. |
||
19 | * Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server. |
||
20 | |||
21 | 1 | Anchi Cheng | h2. Testing |
22 | 7 | Anchi Cheng | |
23 | 9 | Anchi Cheng | h2. Testing with pyscope |
24 | 1 | Anchi Cheng | |
25 | 10 | Anchi Cheng | {{include(DE-12 Installation Pyscope Test)}} |
26 | 1 | Anchi Cheng | |
27 | 10 | Anchi Cheng | h2. Programs to open before Leginon Client: DE Server |
28 | |||
29 | See how to use DE camera usage in |
||
30 | * [[DE-12 introduction]] Please read this!!! |
||
31 | * [[Acquiring summed frame image thru Leginon]] |
||
32 | * [[DDD raw frame saving|Triggering raw frame saving]] |
||
33 | * [[DDD raw frame file transfer|Transferring raw frame images]] off the solid-state drive of DE-12 during data collection to a network drive. |
||
34 | * [[Make DE-12 movie stack|Compiling corrected movie stack]] with Appion using bright and dark images acquired in Leginon |