Actions
Adding DE-12 to Leginon » History » Revision 9
« Previous |
Revision 9/16
(diff)
| Next »
Anchi Cheng, 01/09/2014 09:20 PM
Installation Guide for DE-12 or DE-20¶
DE-12 is on a computer separated from the microscope¶
Please read Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers first.
Extra Package and Installation¶
None.
instruments.cfg¶
DE12 and DE Survey camera are retract/extend together, with the survey camera at off-axis. Therefore zplane should be set at the same value
For example,
[camera1] class: de.DE12 zplane: 50 [camera2] class: de.DESurvey zplane: 50
height and width should be set according to the model.
Setup¶
- Create the folder to store the dose fractionation raw frames. Default location is D:\\frames.
- Setup raw frame file transfer from a network data server.
Testing¶
Testing with pyscope¶
DE-12 Installation Pyscope Test
Programs to open before Leginon Client: Server¶
Updated by Anchi Cheng almost 11 years ago · 9 revisions