Anchi Cheng, 01/09/2014 09:20 PM


Installation Guide for DE-12 or DE-20

DE-12 is on a computer separated from the microscope

Please read Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers first.

Extra Package and Installation

None.

instruments.cfg

DE12 and DE Survey camera are retract/extend together, with the survey camera at off-axis. Therefore zplane should be set at the same value
For example,

[camera1]
class: de.DE12
zplane: 50

[camera2]
class: de.DESurvey
zplane: 50


height and width should be set according to the model.

Setup

  • Create the folder to store the dose fractionation raw frames. Default location is D:\\frames.
  • Setup raw frame file transfer from a network data server.

Testing

Testing with pyscope

DE-12 Installation Pyscope Test

Programs to open before Leginon Client: Server