Project

General

Profile

Adding DE-12 to Leginon » History » Version 13

Anchi Cheng, 02/02/2016 02:30 PM

1 9 Anchi Cheng
h1. Installation Guide for DE-12 or DE-20
2 1 Anchi Cheng
3 9 Anchi Cheng
h2.  DE-12 is on a computer separated from the microscope
4 1 Anchi Cheng
5 9 Anchi Cheng
Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first.
6 1 Anchi Cheng
7 9 Anchi Cheng
h2. Extra Package and Installation
8
9 12 Anchi Cheng
|_.program/package|_.notes|
10
|DirectElectronAPI.exe|(Should come with the camera)|
11
|"Google protobuf":http://code.google.com/p/protobuf/||
12
13
{{include(portobuf installation instruction)}}
14
15
16 9 Anchi Cheng
17
h2. instruments.cfg
18
19
{{include(Instrument.cfg for DE-12)}}
20
height and width should be set according to the model.
21
22
h2. Setup
23
24
* Create the folder to store the dose fractionation raw frames.  Default location is D:\\frames.
25
* Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server.
26
27 1 Anchi Cheng
h2. Testing
28 7 Anchi Cheng
29 9 Anchi Cheng
h2. Testing with pyscope
30 1 Anchi Cheng
31 10 Anchi Cheng
{{include(DE-12 Installation Pyscope Test)}}
32 1 Anchi Cheng
33 10 Anchi Cheng
h2. Programs to open before Leginon Client: DE Server
34
35
See how to use DE camera usage in
36 11 Anchi Cheng
37
* [[DE-12 introduction]] Please read this!!!
38
* [[Acquiring summed frame image thru Leginon]]
39
* [[DDD raw frame saving|Triggering raw frame saving]]
40
* [[DDD raw frame file transfer|Transferring raw frame images]] off the solid-state drive of DE-12 during data collection to a network drive.
41