1 2 Anchi Cheng
h1. AutoIt program, script compilation and testing
2 1 Anchi Cheng
3
h2. Download AutoIt Scripting program on microscope computer
4
5
see https://www.autoitscript.com/site/autoit/downloads/
6
7
h2. Compile *.au3 files
8
9 2 Anchi Cheng
These are found in your_myami/pyscope/autoit/.  Right-click on the file icon should allow you to compile.  Place the resulting exe files at a common place, for example, *C:\Program Files\AutoIt3\* and remember its location to be added to fei.cfg
10 1 Anchi Cheng
11 2 Anchi Cheng
h2. Add the locations of the exe files to fei.cfg of your leginon installation if you have not done so.
12 1 Anchi Cheng
13
# make a copy of your_myami/pyscope/fei.cfg.template and call it fei_temp.cfg
14
# compare fei_temp.cfg with your current fei.cfg used for leginon and modify fei_temp.cfg according to your current configuration.
15
# find the entries related to AUTOIT and fill in the location of your autoit exe files.
16
17
h2. Testing
18
19 2 Anchi Cheng
h3. Beamstop
20
21
# You can double click on BeamstopIn.exe and Beamstopout.exe to see if it does what it should be doing. 
22
23
h3. TUIAcquire
24
25
# Go to TUI interface and CCD/TV Camera panel.
26
# Select BM-Ceta
27
# Double click on TuiAcquire.exe to execute.  This is most likely to fail in some of the validation steps.  Make correction to these settings and try again.
28
# When all validations are passed, you should see TIA starts acquiring movie series using Acquire preset of the CCD/TV Camera panel.
29
# Double click on TuiAcquire.exe again to stop movie series acquisition.
30
31
h3. TiaExportSeries
32
33
# Select TEM Imaging & Analysis (TIA) Window
34
# Make sure Export Series Shortcut is shown like in this screenshot !TIAexport.png!