h1. AutoIt program, script compilation and testing 

 h2. Download AutoIt Scripting program on microscope computer 

 see https://www.autoitscript.com/site/autoit/downloads/ 

 h2. Compile *.au3 files 

 These are found in your_myami/pyscope/autoit/.    Right-click on the file icon should allow you to compile.    Place the resulting exe files at a common place, for example, *C:\Program Files\AutoIt3\* and remember its location to be added to fei.cfg 

 h2. Add the locations of the exe files to fei.cfg of your leginon installation if you have not done so. 

 # make a copy of your_myami/pyscope/fei.cfg.template and call it fei_temp.cfg 
 # compare fei_temp.cfg with your current fei.cfg used for leginon and modify fei_temp.cfg according to your current configuration. 
 # find the entries related to AUTOIT and fill in the location of your autoit exe files. 

 h2. Testing 

 h3. Beamstop 

 # You can double click on BeamstopIn.exe and Beamstopout.exe to see if it does what it should be doing.  

 h3. TUIAcquire 

 # Go to TUI interface and CCD/TV Camera panel. 
 # Select BM-Ceta 
 # Double click on TuiAcquire.exe to execute.    This is most likely to fail in some of the validation steps.    Make correction to these settings and try again. 
 # When all validations are passed, you should see TIA starts acquiring movie series using Acquire preset of the CCD/TV Camera panel. 
 # Double click on TuiAcquire.exe again to stop movie series acquisition. 

 

 h3. TiaExportSeries 

 # Select TEM Imaging & Analysis (TIA) Window that has the series movie acquired from TuiAcquire.exe test above. 
 # Make sure Export Series Shortcut is expanded shown like in this screenshot 
 ## !TIAexport.png! 
 # Double click TiaExportSeries.exe 

 The script should save the series as C:\Users\supervisor\Documents\FEI\TIA\Exported Data\test_t_*.bin 

 AutoIt Script clicks Export Series shortcut icon at a relative position to the corner of the window. If you observe that the position is off, please edit the script at the line 
 <pre> 
    Local $aExportPos[2] = [47, 273] 
 </pre> 
 You can use AutoIt Window Info (part of AutoIt program installation) to find the coordinates that needs to replace 47, 273 (col, row).