Beam Tilt Calibrator » History » Version 1
Amber Herold, 04/23/2010 09:52 AM
1 | 1 | Amber Herold | h1. Beam Tilt Calibrator |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | The Beam Tilt Calibrator is in fact the calibrator for aufocus. This calibration functions |
||
7 | well if the microscope has been well aligned first. The three main functions are defocus |
||
8 | calibration, stigmator calibration and storage/retrieval of eucentric focus |
||
9 | |||
10 | <highlights> |
||
11 | |||
12 | Required bindings to use preset instrument configuration set by presets manager |
||
13 | |||
14 | <simplelist type="vert"> |
||
15 | <member> PresetsManagerNode - (PresetChangedEvent) -> BeamTiltCalibratorNode</member> |
||
16 | </simplelist> |
||
17 | </highlights> |
||
18 | |||
19 | |||
20 | |||
21 | |||
22 | h2. Toolbar ( Leginon/Beam Tilt Calibrator/Toolbar> ) |
||
23 | |||
24 | |||
25 | |||
26 | * Settings = Choose correlation type (phase or cross) and the CCD settings. |
||
27 | |||
28 | |||
29 | * Acquire Image |
||
30 | |||
31 | |||
32 | * Defocus | Stigmators = Choices of types of calibrations to perform. |
||
33 | |||
34 | |||
35 | * Parameter Settings = paramters specific for calibrating the beam tilt |
||
36 | |||
37 | |||
38 | * Calibrate = Begin calibrating |
||
39 | |||
40 | |||
41 | * Abort = Stop the calibration processing |
||
42 | |||
43 | |||
44 | * Measure = Test the beam tilt calibrations |
||
45 | |||
46 | |||
47 | * Eucentric Focus From Scope = store the eucentric focus from the current microscope |
||
48 | settings |
||
49 | |||
50 | |||
51 | * Eucentric Focus To Scope = send the stored Leginon eucentric focus setting to the |
||
52 | microscope |
||
53 | |||
54 | |||
55 | * Rotation Center From Scope = store the rotation center beam tilt value from the |
||
56 | current microscope settings |
||
57 | |||
58 | |||
59 | * Rotation Center To Scope = send the stored Leginon rotation center beam tilt value |
||
60 | to the microscope |
||
61 | |||
62 | |||
63 | * Align Rotation Center = use the current autofocus calibration to align the rotation |
||
64 | center beam tilt. |
||
65 | |||
66 | |||
67 | * Edit current calibration |
||
68 | |||
69 | |||
70 | |||
71 | |||
72 | |||
73 | |||
74 | |||
75 | h2. Settings ( Leginon/Beam Tilt Calibrator/Settings> ) |
||
76 | |||
77 | |||
78 | |||
79 | * Use "phase | cross" correlation = choose the correlation technique that is used to |
||
80 | calibrate the beam tilt |
||
81 | |||
82 | |||
83 | * Override Preset = use this node's <link linkend="camconfig">Instrument and Camera |
||
84 | Configuration</link> to acquire the images in this node |
||
85 | |||
86 | |||
87 | * Correct image for tilt=Apply cosine stretching to the image before correlating the |
||
88 | images |
||
89 | |||
90 | |||
91 | * Settling time = Wait between each beam tilt, used for testing |
||
92 | |||
93 | |||
94 | |||
95 | |||
96 | |||
97 | |||
98 | |||
99 | h2. Defocus or Stigmators |
||
100 | |||
101 | |||
102 | |||
103 | * Defocus = when chosen, the beam dilt will be calibrated in order to correct the |
||
104 | defocus |
||
105 | |||
106 | |||
107 | * Stigmators = when chosen, the beam dilt will be calibrated in order to correct the |
||
108 | stigmation |
||
109 | |||
110 | |||
111 | |||
112 | |||
113 | |||
114 | |||
115 | |||
116 | |||
117 | |||
118 | h2. Parameter Settings |
||
119 | |||
120 | |||
121 | |||
122 | * Beam tilt = expressed in radian, the value gives tilt from center of the the tilt |
||
123 | in each direction, typically 0.01. A higher number means the beam will have a larger |
||
124 | tilt. Too large of a tilt will cause the beam to leave the CCD imaging area. |
||
125 | |||
126 | |||
127 | * First defocus = in meters, typically -2e-6 m (for 50-60k x). A standard defocus |
||
128 | used within the autofocus routine. |
||
129 | |||
130 | This value gives the first defocus from current defocus=0 point where images are |
||
131 | acquired and correlations compared. |
||
132 | |||
133 | |||
134 | * Second defocus = in meters, typically -4e-6 m ( for 50-60k x). A second standard |
||
135 | defocus used within the autofocus routine. |
||
136 | |||
137 | This value gives the second defocus from current defocus=0 point where images are |
||
138 | acquired and correlations compared. |
||
139 | |||
140 | |||
141 | |||
142 | |||
143 | |||
144 | |||
145 | |||
146 | |||
147 | |||
148 | h2. Measure |
||
149 | |||
150 | |||
151 | |||
152 | * Beam tilt = in radian. Typically the same beam tilt used to complete the |
||
153 | calibration (in Parameter Settings) |
||
154 | |||
155 | |||
156 | * Measure = Click this button to test the defocus|stigmator calibrations |
||
157 | |||
158 | The defocus or stigmation that is measured will appear in the corresponding fields |
||
159 | once the measurement has finished. |
||
160 | |||
161 | |||
162 | * Correct Defocus | Correct Stigmator = Send the measured defocus | stigmator |
||
163 | correction values to the microscope. |
||
164 | |||
165 | |||
166 | * Reset Defocus = use this to reset the zero defocus value at the microscope |
||
167 | |||
168 | |||
169 | |||
170 | |||
171 | |||
172 | |||
173 | |||
174 | |||
175 | |||
176 | h2. Eucentric Focus Retrieve from /Send to scope |
||
177 | |||
178 | |||
179 | |||
180 | * Encentric focus value needs to be saved in the database at the magnification and |
||
181 | high tension where stage Z correction is to be made through autofocus routine. |
||
182 | |||
183 | |||
184 | |||
185 | |||
186 | |||
187 | |||
188 | |||
189 | h2. Rotation center Retrieve from /Send to scope |
||
190 | |||
191 | |||
192 | |||
193 | * Rotation center beam tilt value needs to be saved in the database at the |
||
194 | magnification and high tension where stage alpha-tilt distortion correction is to be |
||
195 | made in the focuser node. |
||
196 | |||
197 | |||
198 | |||
199 | |||
200 | |||
201 | |||
202 | |||
203 | h2. Align Rotation Center |
||
204 | |||
205 | |||
206 | |||