Beam Tilt Calibrator » History » Version 1

Amber Herold, 04/23/2010 09:52 AM

1 1 Amber Herold
h1. Beam Tilt Calibrator
2
3
4
5
6
The Beam Tilt Calibrator is in fact the calibrator for aufocus. This calibration functions
7
well if the microscope has been well aligned first. The three main functions are defocus
8
calibration, stigmator calibration and storage/retrieval of eucentric focus
9
10
<highlights>
11
12
Required bindings to use preset instrument configuration set by presets manager
13
14
<simplelist type="vert">
15
<member> PresetsManagerNode - (PresetChangedEvent) -> BeamTiltCalibratorNode</member>
16
</simplelist>
17
</highlights>
18
19
20
21
22
h2.  Toolbar ( Leginon/Beam Tilt Calibrator/Toolbar> )
23
24
25
26
*  Settings = Choose correlation type (phase or cross) and the CCD settings.
27
28
29
*  Acquire Image
30
31
32
*  Defocus | Stigmators = Choices of types of calibrations to perform.
33
34
35
*  Parameter Settings = paramters specific for calibrating the beam tilt
36
37
38
*  Calibrate = Begin calibrating
39
40
41
*  Abort = Stop the calibration processing
42
43
44
*  Measure = Test the beam tilt calibrations
45
46
47
*  Eucentric Focus From Scope = store the eucentric focus from the current microscope
48
settings
49
50
51
*  Eucentric Focus To Scope = send the stored Leginon eucentric focus setting to the
52
microscope
53
54
55
*  Rotation Center From Scope = store the rotation center beam tilt value from the
56
current microscope settings
57
58
59
*  Rotation Center To Scope = send the stored Leginon rotation center beam tilt value
60
to the microscope
61
62
63
*  Align Rotation Center = use the current autofocus calibration to align the rotation
64
center beam tilt.
65
66
67
*  Edit current calibration
68
69
70
71
72
73
74
75
h2.  Settings ( Leginon/Beam Tilt Calibrator/Settings> )
76
77
78
79
*  Use "phase | cross" correlation = choose the correlation technique that is used to
80
calibrate the beam tilt
81
82
83
*  Override Preset = use this node's <link linkend="camconfig">Instrument and Camera
84
Configuration</link> to acquire the images in this node
85
86
87
*  Correct image for tilt=Apply cosine stretching to the image before correlating the
88
images
89
90
91
*  Settling time = Wait between each beam tilt, used for testing
92
93
94
95
96
97
98
99
h2.  Defocus or Stigmators
100
101
102
103
*  Defocus = when chosen, the beam dilt will be calibrated in order to correct the
104
defocus
105
106
107
*  Stigmators = when chosen, the beam dilt will be calibrated in order to correct the
108
stigmation
109
110
111
112
113
114
115
116
117
118
h2.  Parameter Settings
119
120
121
122
*  Beam tilt = expressed in radian, the value gives tilt from center of the the tilt
123
in each direction, typically 0.01. A higher number means the beam will have a larger
124
tilt. Too large of a tilt will cause the beam to leave the CCD imaging area.
125
126
127
*  First defocus = in meters, typically -2e-6 m (for 50-60k x). A standard defocus
128
used within the autofocus routine.
129
130
This value gives the first defocus from current defocus=0 point where images are
131
acquired and correlations compared.
132
133
134
*  Second defocus = in meters, typically -4e-6 m ( for 50-60k x). A second standard
135
defocus used within the autofocus routine.
136
137
This value gives the second defocus from current defocus=0 point where images are
138
acquired and correlations compared.
139
140
141
142
143
144
145
146
147
148
h2.  Measure
149
150
151
152
*  Beam tilt = in radian. Typically the same beam tilt used to complete the
153
calibration (in Parameter Settings)
154
155
156
*  Measure = Click this button to test the defocus|stigmator calibrations
157
158
The defocus or stigmation that is measured will appear in the corresponding fields
159
once the measurement has finished.
160
161
162
*  Correct Defocus | Correct Stigmator = Send the measured defocus | stigmator
163
correction values to the microscope.
164
165
166
*  Reset Defocus = use this to reset the zero defocus value at the microscope
167
168
169
170
171
172
173
174
175
176
h2.  Eucentric Focus Retrieve from /Send to scope
177
178
179
180
*  Encentric focus value needs to be saved in the database at the magnification and
181
high tension where stage Z correction is to be made through autofocus routine.
182
183
184
185
186
187
188
189
h2.  Rotation center Retrieve from /Send to scope
190
191
192
193
*  Rotation center beam tilt value needs to be saved in the database at the
194
magnification and high tension where stage alpha-tilt distortion correction is to be
195
made in the focuser node.
196
197
198
199
200
201
202
203
h2.  Align Rotation Center
204
205
206
207
*  Align the beam tilt to the rotation center based on the current defocus/stigmator
208
calibration.