| 4 | |||
| 5 | |||
| 6 | The Beam Tilt Calibrator is in fact the calibrator for aufocus. This calibration functions | ||
| 7 | well if the microscope has been well aligned first. The three main functions are defocus | ||
| 8 | calibration, stigmator calibration and storage/retrieval of eucentric focus | ||
| 9 | |||
| 10 | <highlights> | ||
| 11 | |||
| 12 | Required bindings to use preset instrument configuration set by presets manager | ||
| 13 | |||
| 14 | <simplelist type="vert"> | ||
| 15 | <member> PresetsManagerNode - (PresetChangedEvent) -> BeamTiltCalibratorNode</member> | ||
| 16 | </simplelist> | ||
| 17 | </highlights> | ||
| 18 | |||
| 19 | |||
| 20 | |||
| 21 | |||
| 22 | h2. Toolbar ( Leginon/Beam Tilt Calibrator/Toolbar> ) | ||
| 23 | |||
| 24 | |||
| 25 | |||
| 26 | * Settings = Choose correlation type (phase or cross) and the CCD settings. | ||
| 27 | |||
| 28 | |||
| 29 | * Acquire Image | ||
| 30 | |||
| 31 | |||
| 32 | * Defocus | Stigmators = Choices of types of calibrations to perform. | ||
| 33 | |||
| 34 | |||
| 35 | * Parameter Settings = paramters specific for calibrating the beam tilt | ||
| 36 | |||
| 37 | |||
| 38 | * Calibrate = Begin calibrating | ||
| 39 | |||
| 40 | |||
| 41 | * Abort = Stop the calibration processing | ||
| 42 | |||
| 43 | |||
| 44 | * Measure = Test the beam tilt calibrations | ||
| 45 | |||
| 46 | |||
| 47 | * Eucentric Focus From Scope = store the eucentric focus from the current microscope | ||
| 48 | settings | ||
| 49 | |||
| 50 | |||
| 51 | * Eucentric Focus To Scope = send the stored Leginon eucentric focus setting to the | ||
| 52 | microscope | ||
| 53 | |||
| 54 | |||
| 55 | * Rotation Center From Scope = store the rotation center beam tilt value from the | ||
| 56 | current microscope settings | ||
| 57 | |||
| 58 | |||
| 59 | * Rotation Center To Scope = send the stored Leginon rotation center beam tilt value | ||
| 60 | to the microscope | ||
| 61 | |||
| 62 | |||
| 63 | * Align Rotation Center = use the current autofocus calibration to align the rotation | ||
| 64 | center beam tilt. | ||
| 65 | |||
| 66 | |||
| 67 | * Edit current calibration | ||
| 68 | |||
| 69 | |||
| 70 | |||
| 71 | |||
| 72 | |||
| 73 | |||
| 74 | |||
| 75 | h2. Settings ( Leginon/Beam Tilt Calibrator/Settings> ) | ||
| 76 | |||
| 77 | |||
| 78 | |||
| 79 | * Use "phase | cross" correlation = choose the correlation technique that is used to | ||
| 80 | calibrate the beam tilt | ||
| 81 | |||
| 82 | |||
| 83 | * Override Preset = use this node's <link linkend="camconfig">Instrument and Camera | ||
| 84 | Configuration</link> to acquire the images in this node | ||
| 85 | |||
| 86 | |||
| 87 | * Correct image for tilt=Apply cosine stretching to the image before correlating the | ||
| 88 | images | ||
| 89 | |||
| 90 | |||
| 91 | * Settling time = Wait between each beam tilt, used for testing | ||
| 92 | |||
| 93 | |||
| 94 | |||
| 95 | |||
| 96 | |||
| 97 | |||
| 98 | |||
| 99 | h2. Defocus or Stigmators | ||
| 100 | |||
| 101 | |||
| 102 | |||
| 103 | * Defocus = when chosen, the beam dilt will be calibrated in order to correct the | ||
| 104 | defocus | ||
| 105 | |||
| 106 | |||
| 107 | * Stigmators = when chosen, the beam dilt will be calibrated in order to correct the | ||
| 108 | stigmation | ||
| 109 | |||
| 110 | |||
| 111 | |||
| 112 | |||
| 113 | |||
| 114 | |||
| 115 | |||
| 116 | |||
| 117 | |||
| 118 | h2. Parameter Settings | ||
| 119 | |||
| 120 | |||
| 121 | |||
| 122 | * Beam tilt = expressed in radian, the value gives tilt from center of the the tilt | ||
| 123 | in each direction, typically 0.01. A higher number means the beam will have a larger | ||
| 124 | tilt. Too large of a tilt will cause the beam to leave the CCD imaging area. | ||
| 125 | |||
| 126 | |||
| 127 | * First defocus = in meters, typically -2e-6 m (for 50-60k x). A standard defocus | ||
| 128 | used within the autofocus routine. | ||
| 129 | |||
| 130 | This value gives the first defocus from current defocus=0 point where images are | ||
| 131 | acquired and correlations compared. | ||
| 132 | |||
| 133 | |||
| 134 | * Second defocus = in meters, typically -4e-6 m ( for 50-60k x). A second standard | ||
| 135 | defocus used within the autofocus routine. | ||
| 136 | |||
| 137 | This value gives the second defocus from current defocus=0 point where images are | ||
| 138 | acquired and correlations compared. | ||
| 139 | |||
| 140 | |||
| 141 | |||
| 142 | |||
| 143 | |||
| 144 | |||
| 145 | |||
| 146 | |||
| 147 | |||
| 148 | h2. Measure | ||
| 149 | |||
| 150 | |||
| 151 | |||
| 152 | * Beam tilt = in radian. Typically the same beam tilt used to complete the | ||
| 153 | calibration (in Parameter Settings) | ||
| 154 | |||
| 155 | |||
| 156 | * Measure = Click this button to test the defocus|stigmator calibrations | ||
| 157 | |||
| 158 | The defocus or stigmation that is measured will appear in the corresponding fields | ||
| 159 | once the measurement has finished. | ||
| 160 | |||
| 161 | |||
| 162 | * Correct Defocus | Correct Stigmator = Send the measured defocus | stigmator | ||
| 163 | correction values to the microscope. | ||
| 164 | |||
| 165 | |||
| 166 | * Reset Defocus = use this to reset the zero defocus value at the microscope | ||
| 167 | |||
| 168 | |||
| 169 | |||
| 170 | |||
| 171 | |||
| 172 | |||
| 173 | |||
| 174 | |||
| 175 | |||
| 176 | h2. Eucentric Focus Retrieve from /Send to scope | ||
| 177 | |||
| 178 | |||
| 179 | |||
| 180 | * Encentric focus value needs to be saved in the database at the magnification and | ||
| 181 | high tension where stage Z correction is to be made through autofocus routine. | ||
| 182 | |||
| 183 | |||
| 184 | |||
| 185 | |||
| 186 | |||
| 187 | |||
| 188 | |||
| 189 | h2. Rotation center Retrieve from /Send to scope | ||
| 190 | |||
| 191 | |||
| 192 | |||
| 193 | * Rotation center beam tilt value needs to be saved in the database at the | ||
| 194 | magnification and high tension where stage alpha-tilt distortion correction is to be | ||
| 195 | made in the focuser node. | ||
| 196 | |||
| 197 | |||
| 198 | |||
| 199 | |||
| 200 | |||
| 201 | |||
| 202 | |||
| 203 | h2. Align Rotation Center | ||
| 204 | |||
| 205 | |||
| 206 | |||
| 207 | * Align the beam tilt to the rotation center based on the current defocus/stigmator | ||
| 208 | calibration. | ||
| 209 | 2 | Amber Herold | |
| 210 | |||
| 211 | ______ | ||
| 212 | |||
| 213 | [[BeamFixer|< BeamFixer]] | [[Beam Tilt Imager|Beam Tilt Imager >]] | ||
| 214 | |||
| 215 | ______ |