h1. Beam Tilt Calibrator 




 The Beam Tilt Calibrator is in fact the calibrator for aufocus. This calibration functions 
 well if the microscope has been well aligned first. The three main functions are defocus 
 calibration, stigmator calibration and storage/retrieval of eucentric focus 

 <highlights> 

 Required bindings to use preset instrument configuration set by presets manager 

 <simplelist type="vert"> 
 <member> PresetsManagerNode - (PresetChangedEvent) -> BeamTiltCalibratorNode</member> 
 </simplelist> 
 </highlights> 




 h2.    Toolbar ( Leginon/Beam Tilt Calibrator/Toolbar> ) 



 *    Settings = Choose correlation type (phase or cross) and the CCD settings. 


 *    Acquire Image 


 *    Defocus | Stigmators = Choices of types of calibrations to perform. 


 *    Parameter Settings = paramters specific for calibrating the beam tilt 


 *    Calibrate = Begin calibrating 


 *    Abort = Stop the calibration processing 


 *    Measure = Test the beam tilt calibrations 


 *    Eucentric Focus From Scope = store the eucentric focus from the current microscope 
 settings 


 *    Eucentric Focus To Scope = send the stored Leginon eucentric focus setting to the 
 microscope 


 *    Rotation Center From Scope = store the rotation center beam tilt value from the 
 current microscope settings 


 *    Rotation Center To Scope = send the stored Leginon rotation center beam tilt value 
 to the microscope 


 *    Align Rotation Center = use the current autofocus calibration to align the rotation 
 center beam tilt. 


 *    Edit current calibration 







 h2.    Settings ( Leginon/Beam Tilt Calibrator/Settings> ) 



 *    Use "phase | cross" correlation = choose the correlation technique that is used to 
 calibrate the beam tilt 


 *    Override Preset = use this node's <link linkend="camconfig">Instrument and Camera 
 Configuration</link> to acquire the images in this node 


 *    Correct image for tilt=Apply cosine stretching to the image before correlating the 
 images 


 *    Settling time = Wait between each beam tilt, used for testing 







 h2.    Defocus or Stigmators 



 *    Defocus = when chosen, the beam dilt will be calibrated in order to correct the 
 defocus 


 *    Stigmators = when chosen, the beam dilt will be calibrated in order to correct the 
 stigmation 









 h2.    Parameter Settings 



 *    Beam tilt = expressed in radian, the value gives tilt from center of the the tilt 
 in each direction, typically 0.01. A higher number means the beam will have a larger 
 tilt. Too large of a tilt will cause the beam to leave the CCD imaging area. 


 *    First defocus = in meters, typically -2e-6 m (for 50-60k x). A standard defocus 
 used within the autofocus routine. 

 This value gives the first defocus from current defocus=0 point where images are 
 acquired and correlations compared. 


 *    Second defocus = in meters, typically -4e-6 m ( for 50-60k x). A second standard 
 defocus used within the autofocus routine. 

 This value gives the second defocus from current defocus=0 point where images are 
 acquired and correlations compared. 









 h2.    Measure 



 *    Beam tilt = in radian. Typically the same beam tilt used to complete the 
 calibration (in Parameter Settings) 


 *    Measure = Click this button to test the defocus|stigmator calibrations 

 The defocus or stigmation that is measured will appear in the corresponding fields 
 once the measurement has finished. 


 *    Correct Defocus | Correct Stigmator = Send the measured defocus | stigmator 
 correction values to the microscope. 


 *    Reset Defocus = use this to reset the zero defocus value at the microscope 









 h2.    Eucentric Focus Retrieve from /Send to scope 



 *    Encentric focus value needs to be saved in the database at the magnification and 
 high tension where stage Z correction is to be made through autofocus routine. 







 h2.    Rotation center Retrieve from /Send to scope 



 *    Rotation center beam tilt value needs to be saved in the database at the 
 magnification and high tension where stage alpha-tilt distortion correction is to be 
 made in the focuser node. 







 h2.    Align Rotation Center 



 *    Align the beam tilt to the rotation center based on the current defocus/stigmator 
 calibration. 


 ______ 

 [[BeamFixer|< BeamFixer]] | [[Beam Tilt Imager|Beam Tilt Imager >]] 