Beam Tilt Calibrator » History » Version 3
Eric Hou, 06/24/2010 04:36 PM
1 | 1 | Amber Herold | h1. Beam Tilt Calibrator |
---|---|---|---|
2 | |||
3 | 3 | Eric Hou | The Beam Tilt Calibrator is in fact the calibrator for aufocus. This calibration functions well if the microscope has been well aligned first. The three main functions are defocus calibration, stigmator calibration and storage/retrieval of eucentric focus |
4 | 1 | Amber Herold | |
5 | Required bindings to use preset instrument configuration set by presets manager |
||
6 | |||
7 | 3 | Eric Hou | PresetsManagerNode - (PresetChangedEvent) -> BeamTiltCalibratorNode |
8 | 1 | Amber Herold | |
9 | h2. Toolbar ( Leginon/Beam Tilt Calibrator/Toolbar> ) |
||
10 | |||
11 | * Settings = Choose correlation type (phase or cross) and the CCD settings. |
||
12 | |||
13 | * Acquire Image |
||
14 | |||
15 | * Defocus | Stigmators = Choices of types of calibrations to perform. |
||
16 | |||
17 | * Parameter Settings = paramters specific for calibrating the beam tilt |
||
18 | |||
19 | * Calibrate = Begin calibrating |
||
20 | |||
21 | * Abort = Stop the calibration processing |
||
22 | |||
23 | * Measure = Test the beam tilt calibrations |
||
24 | |||
25 | 3 | Eric Hou | * Eucentric Focus From Scope = store the eucentric focus from the current microscope settings |
26 | 1 | Amber Herold | |
27 | 3 | Eric Hou | * Eucentric Focus To Scope = send the stored Leginon eucentric focus setting to the microscope |
28 | 1 | Amber Herold | |
29 | 3 | Eric Hou | * Rotation Center From Scope = store the rotation center beam tilt value from the current microscope settings |
30 | 1 | Amber Herold | |
31 | 3 | Eric Hou | * Rotation Center To Scope = send the stored Leginon rotation center beam tilt value to the microscope |
32 | 1 | Amber Herold | |
33 | 3 | Eric Hou | * Align Rotation Center = use the current autofocus calibration to align the rotation center beam tilt. |
34 | 1 | Amber Herold | |
35 | * Edit current calibration |
||
36 | |||
37 | h2. Settings ( Leginon/Beam Tilt Calibrator/Settings> ) |
||
38 | |||
39 | * Use "phase | cross" correlation = choose the correlation technique that is used to |
||
40 | calibrate the beam tilt |
||
41 | |||
42 | 3 | Eric Hou | * Override Preset = use this node's <link linkend="camconfig">Instrument and Camera Configuration</link> to acquire the images in this node |
43 | 1 | Amber Herold | |
44 | 3 | Eric Hou | * Correct image for tilt=Apply cosine stretching to the image before correlating the images |
45 | 1 | Amber Herold | |
46 | * Settling time = Wait between each beam tilt, used for testing |
||
47 | |||
48 | h2. Defocus or Stigmators |
||
49 | |||
50 | 3 | Eric Hou | * Defocus = when chosen, the beam dilt will be calibrated in order to correct the defocus |
51 | 1 | Amber Herold | |
52 | 3 | Eric Hou | * Stigmators = when chosen, the beam dilt will be calibrated in order to correct the stigmation |
53 | 1 | Amber Herold | |
54 | h2. Parameter Settings |
||
55 | |||
56 | 3 | Eric Hou | * Beam tilt = expressed in radian, the value gives tilt from center of the the tilt in each direction, typically 0.01. A higher number means the beam will have a larger tilt. Too large of a tilt will cause the beam to leave the CCD imaging area. |
57 | 1 | Amber Herold | |
58 | 3 | Eric Hou | * First defocus = in meters, typically -2e-6 m (for 50-60k x). A standard defocus used within the autofocus routine. |
59 | 1 | Amber Herold | |
60 | 3 | Eric Hou | This value gives the first defocus from current defocus=0 point where images are acquired and correlations compared. |
61 | 1 | Amber Herold | |
62 | 3 | Eric Hou | * Second defocus = in meters, typically -4e-6 m ( for 50-60k x). A second standard defocus used within the autofocus routine. |
63 | 1 | Amber Herold | |
64 | 3 | Eric Hou | This value gives the second defocus from current defocus=0 point where images are acquired and correlations compared. |
65 | 1 | Amber Herold | |
66 | h2. Measure |
||
67 | |||
68 | 3 | Eric Hou | * Beam tilt = in radian. Typically the same beam tilt used to complete the calibration (in Parameter Settings) |
69 | 1 | Amber Herold | |
70 | * Measure = Click this button to test the defocus|stigmator calibrations |
||
71 | |||
72 | 3 | Eric Hou | The defocus or stigmation that is measured will appear in the corresponding fields once the measurement has finished. |
73 | 1 | Amber Herold | |
74 | 3 | Eric Hou | * Correct Defocus | Correct Stigmator = Send the measured defocus | stigmator correction values to the microscope. |
75 | 1 | Amber Herold | |
76 | * Reset Defocus = use this to reset the zero defocus value at the microscope |
||
77 | |||
78 | h2. Eucentric Focus Retrieve from /Send to scope |
||
79 | |||
80 | 3 | Eric Hou | * Encentric focus value needs to be saved in the database at the magnification and high tension where stage Z correction is to be made through autofocus routine. |
81 | 1 | Amber Herold | |
82 | h2. Rotation center Retrieve from /Send to scope |
||
83 | |||
84 | 3 | Eric Hou | * Rotation center beam tilt value needs to be saved in the database at the magnification and high tension where stage alpha-tilt distortion correction is to be made in the focuser node. |
85 | 1 | Amber Herold | |
86 | h2. Align Rotation Center |
||
87 | |||
88 | 3 | Eric Hou | * Align the beam tilt to the rotation center based on the current defocus/stigmator calibration. |
89 | 2 | Amber Herold | |
90 | ______ |
||
91 | |||
92 | [[BeamFixer|< BeamFixer]] | [[Beam Tilt Imager|Beam Tilt Imager >]] |
||