Project

General

Profile

Beam Tilt Calibrator » History » Version 4

Eric Hou, 06/24/2010 04:37 PM

1 1 Amber Herold
h1. Beam Tilt Calibrator
2
3 3 Eric Hou
The Beam Tilt Calibrator is in fact the calibrator for aufocus. This calibration functions well if the microscope has been well aligned first. The three main functions are defocus calibration, stigmator calibration and storage/retrieval of eucentric focus
4 1 Amber Herold
5
Required bindings to use preset instrument configuration set by presets manager
6
7 3 Eric Hou
PresetsManagerNode - (PresetChangedEvent) -> BeamTiltCalibratorNode
8 1 Amber Herold
9
h2.  Toolbar ( Leginon/Beam Tilt Calibrator/Toolbar> )
10
11
*  Settings = Choose correlation type (phase or cross) and the CCD settings.
12
13
*  Acquire Image
14
15
*  Defocus | Stigmators = Choices of types of calibrations to perform.
16
17
*  Parameter Settings = paramters specific for calibrating the beam tilt
18
19
*  Calibrate = Begin calibrating
20
21
*  Abort = Stop the calibration processing
22
23
*  Measure = Test the beam tilt calibrations
24
25 3 Eric Hou
*  Eucentric Focus From Scope = store the eucentric focus from the current microscope settings
26 1 Amber Herold
27 3 Eric Hou
*  Eucentric Focus To Scope = send the stored Leginon eucentric focus setting to the microscope
28 1 Amber Herold
29 3 Eric Hou
*  Rotation Center From Scope = store the rotation center beam tilt value from the current microscope settings
30 1 Amber Herold
31 3 Eric Hou
*  Rotation Center To Scope = send the stored Leginon rotation center beam tilt value to the microscope
32 1 Amber Herold
33 3 Eric Hou
*  Align Rotation Center = use the current autofocus calibration to align the rotation center beam tilt.
34 1 Amber Herold
35
*  Edit current calibration
36
37
h2.  Settings ( Leginon/Beam Tilt Calibrator/Settings> )
38
39
*  Use "phase | cross" correlation = choose the correlation technique that is used to
40
calibrate the beam tilt
41
42 3 Eric Hou
*  Override Preset = use this node's <link linkend="camconfig">Instrument and Camera Configuration</link> to acquire the images in this node
43 1 Amber Herold
44 3 Eric Hou
*  Correct image for tilt=Apply cosine stretching to the image before correlating the images
45 1 Amber Herold
46
*  Settling time = Wait between each beam tilt, used for testing
47
48
h2.  Defocus or Stigmators
49
50 3 Eric Hou
*  Defocus = when chosen, the beam dilt will be calibrated in order to correct the defocus
51 1 Amber Herold
52 3 Eric Hou
*  Stigmators = when chosen, the beam dilt will be calibrated in order to correct the stigmation
53 1 Amber Herold
54
h2.  Parameter Settings
55
56 3 Eric Hou
*  Beam tilt = expressed in radian, the value gives tilt from center of the the tilt in each direction, typically 0.01. A higher number means the beam will have a larger tilt. Too large of a tilt will cause the beam to leave the CCD imaging area.
57 1 Amber Herold
58 3 Eric Hou
*  First defocus = in meters, typically -2e-6 m (for 50-60k x). A standard defocus used within the autofocus routine.
59 1 Amber Herold
60 3 Eric Hou
 This value gives the first defocus from current defocus=0 point where images are acquired and correlations compared.
61 1 Amber Herold
62 3 Eric Hou
*  Second defocus = in meters, typically -4e-6 m ( for 50-60k x). A second standard defocus used within the autofocus routine.
63 1 Amber Herold
64 4 Eric Hou
 This value gives the second defocus from current defocus=0 point where images are acquired and correlations compared.
65 1 Amber Herold
66
h2.  Measure
67
68 3 Eric Hou
*  Beam tilt = in radian. Typically the same beam tilt used to complete the calibration (in Parameter Settings)
69 1 Amber Herold
70
*  Measure = Click this button to test the defocus|stigmator calibrations
71
72 3 Eric Hou
 The defocus or stigmation that is measured will appear in the corresponding fields once the measurement has finished.
73 1 Amber Herold
74 3 Eric Hou
*  Correct Defocus | Correct Stigmator = Send the measured defocus | stigmator correction values to the microscope.
75 1 Amber Herold
76
*  Reset Defocus = use this to reset the zero defocus value at the microscope
77
78
h2.  Eucentric Focus Retrieve from /Send to scope
79
80 3 Eric Hou
*  Encentric focus value needs to be saved in the database at the magnification and high tension where stage Z correction is to be made through autofocus routine.
81 1 Amber Herold
82
h2.  Rotation center Retrieve from /Send to scope
83
84 3 Eric Hou
*  Rotation center beam tilt value needs to be saved in the database at the magnification and high tension where stage alpha-tilt distortion correction is to be made in the focuser node.
85 1 Amber Herold
86
h2.  Align Rotation Center
87
88 3 Eric Hou
*  Align the beam tilt to the rotation center based on the current defocus/stigmator calibration.
89 2 Amber Herold
90
______
91
92
[[BeamFixer|< BeamFixer]] | [[Beam Tilt Imager|Beam Tilt Imager >]]
93
94
______