Project

General

Profile

Calibration without presets » History » Version 1

Amber Herold, 04/22/2010 04:49 PM

1 1 Amber Herold
h1. Calibration without presets
2
3
4
Calibrations can be made without presets although not as conveniently. For each of the
5
calibration node, the following need to be done before the calibration or even before
6
acquiring a test image.
7
8
9
10
h2.  Select the TEM and Camera Configuration
11
12
13
14
15
16
*  Leginon/Some Calibraion Node/Toolbar> Click the "Settings" button !http://emg.nysbc.org/software/leginon/images/icons/settings.png!.
17
18
19
*  Leginon/Same Calibration Node/Settings> Override Preset = yes
20
21
22
*  Leginon/Same Calibration Node/Settings/Instrument> TEM = your microscope; CCD
23
Camera = your CCD camera
24
25
26
*  Leginon/Same Calibration Node/Camera Configuration> Choose according to your
27
need. Usually a highly binned configuration that uses the whole camera.
28
29
30
*  Leginon/Same Calibration Node/Camera Configuration> Click OK to save and close
31
the window.
32
33
34
*  Leginon/Pixel Size/Same Calibration Node> Acquire a test image with the camera
35
tool!http://emg.nysbc.org/software/leginon/images/icons/acquire.png! to confirrm that the TEM and camera settings are valid.