2 | |||
3 | |||
4 | Calibrations can be made without presets although not as conveniently. For each of the |
||
5 | calibration node, the following need to be done before the calibration or even before |
||
6 | acquiring a test image. |
||
7 | |||
8 | |||
9 | |||
10 | h2. Select the TEM and Camera Configuration |
||
11 | |||
12 | |||
13 | |||
14 | |||
15 | |||
16 | * Leginon/Some Calibraion Node/Toolbar> Click the "Settings" button !http://emg.nysbc.org/software/leginon/images/icons/settings.png!. |
||
17 | |||
18 | |||
19 | * Leginon/Same Calibration Node/Settings> Override Preset = yes |
||
20 | |||
21 | |||
22 | * Leginon/Same Calibration Node/Settings/Instrument> TEM = your microscope; CCD |
||
23 | Camera = your CCD camera |
||
24 | |||
25 | |||
26 | * Leginon/Same Calibration Node/Camera Configuration> Choose according to your |
||
27 | need. Usually a highly binned configuration that uses the whole camera. |
||
28 | |||
29 | |||
30 | * Leginon/Same Calibration Node/Camera Configuration> Click OK to save and close |
||
31 | the window. |
||
32 | |||
33 | |||
34 | * Leginon/Pixel Size/Same Calibration Node> Acquire a test image with the camera |
||
35 | tool!http://emg.nysbc.org/software/leginon/images/icons/acquire.png! to confirrm that the TEM and camera settings are valid. |
||
36 | 2 | Amber Herold | |
37 | |||
38 | |||
39 | |||
40 | ______ |
||