Calibration without presets » History » Version 3
  Amber Herold, 06/23/2010 12:37 PM 
  
| 1 | 1 | Amber Herold | h1. Calibration without presets | 
|---|---|---|---|
| 2 | |||
| 3 | |||
| 4 | Calibrations can be made without presets although not as conveniently. For each of the | ||
| 5 | calibration node, the following need to be done before the calibration or even before | ||
| 6 | acquiring a test image. | ||
| 7 | |||
| 8 | |||
| 9 | |||
| 10 | h2. Select the TEM and Camera Configuration | ||
| 11 | |||
| 12 | |||
| 13 | |||
| 14 | |||
| 15 | |||
| 16 | * Leginon/Some Calibraion Node/Toolbar> Click the "Settings" button !http://emg.nysbc.org/software/leginon/images/icons/settings.png!. | ||
| 17 | |||
| 18 | |||
| 19 | * Leginon/Same Calibration Node/Settings> Override Preset = yes | ||
| 20 | |||
| 21 | |||
| 22 | * Leginon/Same Calibration Node/Settings/Instrument> TEM = your microscope; CCD | ||
| 23 | Camera = your CCD camera | ||
| 24 | |||
| 25 | |||
| 26 | * Leginon/Same Calibration Node/Camera Configuration> Choose according to your | ||
| 27 | need. Usually a highly binned configuration that uses the whole camera. | ||
| 28 | |||
| 29 | |||
| 30 | * Leginon/Same Calibration Node/Camera Configuration> Click OK to save and close | ||
| 31 | the window. | ||
| 32 | |||
| 33 | |||
| 34 | * Leginon/Pixel Size/Same Calibration Node> Acquire a test image with the camera | ||
| 35 | 3 | Amber Herold | tool !http://emg.nysbc.org/software/leginon/images/icons/acquire.png! to confirm that the TEM and camera settings are valid. | 
| 36 | 2 | Amber Herold | |
| 37 | |||
| 38 | |||
| 39 | |||
| 40 | ______ | ||
| 41 | |||
| 42 | [[Creating Presets for the first time|< Creating Presets for the first time]] | [[Pixel Size Calibration|Pixel Size Calibration >]] | ||