Camera length calibration » History » Version 2
Anchi Cheng, 04/29/2020 05:10 PM
1 | 1 | Anchi Cheng | h1. Camera length calibration |
---|---|---|---|
2 | |||
3 | Camera length calibration can be simply done by acquiring a diffraction image of cross-grating replica and enter the expected d-spacing of Au-Pd power rings and measure the diameter of the ring in Leginon gui. |
||
4 | |||
5 | 2 | Anchi Cheng | * This is only tested on TFS Ceta and Tvips F416 at 2kx2k bin 2 |
6 | |||
7 | # Send a diffraction tem preset. *This should be a square image*. |
||
8 | 1 | Anchi Cheng | # Select "Camera Length" node |
9 | # Click "Acquire Image" |
||
10 | # Measure the diameter of the ring in pixels. |
||
11 | # Click "Ruler" tool to enter the expected d-spacing and measurement, and then click "Calculate" |
||
12 | # Multiple measurements can be made with the same preset and the camera length averaged. |
||
13 | # Select the camera lengths to be averaged and saved, and save them. |
||
14 | |||
15 | Direct Edit and Extrapolation is possible as in Pixel Size Calibrator. |
||
16 |