Actions
Check Cs value match before using Leginon after update » History » Revision 1
Revision 1/6
| Next »
Anchi Cheng, 06/07/2012 01:51 PM
Check Cs value match before using Leginon after update¶
Run the following script on any of the computer where you have done the update on.
On Linux:
$PYTHONSITEPKG/leginon/cscheck.py
On Windows:
C://Python25/Lib/site-packages/leginon/cscheck.py
You should get a list of microscope where you have saved images on in a Leginon session and the Cs values that you entered using schema-r15653.py Check that against the value in instruments.cfg on the named microscope. The two must match. Otherwise Leginon will complain and you may lose all your old calibrations.
< How to Update from v2.1 (Microscope Windows Computer)