Check Cs value match before using Leginon after update

Run the following script on any of the computer where you have done the update on.

On Linux:

$PYTHONSITEPKG/leginon/cscheck.py

On Windows:

C://Python25/Lib/site-packages/leginon/cscheck.py

You should get a list of microscope where you have saved images on in a Leginon session and the Cs values that you entered using schema-r15653.py Check that against the value in instruments.cfg on the named microscope. The two must match. Otherwise Leginon will complain and you may lose all your old calibrations.


< How to Update from v2.1 (Microscope Windows Computer)