Exposure Targeting Set-up for MSI-T for Four holes » History » Version 7
Anchi Cheng, 08/16/2016 03:49 PM
1 | 1 | Anchi Cheng | h1. Exposure Targeting Set-up for MSI-T for Four holes |
---|---|---|---|
2 | |||
3 | A good way to improve the throughput if there are multiple holes visible in a hl image is to target all four holes. |
||
4 | |||
5 | We can set up for 4 holes with this more challenging example: |
||
6 | 2 | Anchi Cheng | |
7 | h2. About this image: |
||
8 | |||
9 | 3 | Anchi Cheng | !original.png! |
10 | 2 | Anchi Cheng | |
11 | h2. Leginon/Exposure Targeting/Template> |
||
12 | |||
13 | 6 | Anchi Cheng | * Cross correlation works better than phase correlation in this case because the hole does not have sharp edge and not perfectly round. |
14 | 5 | Anchi Cheng | * Adjust the value to fill the mean before correlation to minimize the correlation along the edge of grid bar. |
15 | * Invert template for correlation if needed. |
||
16 | |||
17 | 4 | Anchi Cheng | !template_corr_settings.png! !template_corr.png! |
18 | 7 | Anchi Cheng | |
19 | h2. Leginon/Exposure Targeting/Threshold> |
||
20 | |||
21 | * Fairly easy to set. |
||
22 | !threshold_settings.png! !threshold.png! |
||
23 | |||
24 | h2. Leginon/Exposure Targeting/Blobs> |
||
25 | |||
26 | * Allow more blobs than the intended 4. |
||
27 | !blobs_settings! !blobs.png! |
||
28 | |||
29 | h2. Leginon/Exposure Targeting/Lattice> |
||
30 | |||
31 | * Similar to that in Hole Targeting/Lattice step. |
||
32 | ** Make measurement of the lattice distance |
||
33 | ** Estimate the reference intensity for ice thickness filter. |
||
34 | !lattice_settings.png! !threshold_with_lattice_measurement.png! !lattice.png! |
||
35 | |||
36 | h2. Leginon/Exposure Targeting/Acquisition> |
||
37 | |||
38 | * Set up focus template but DO NOT eliminate focus targets using ice thickness criteria. This works because Leginon (3.1 and above) will automatically choose the center most focus target. |
||
39 | !acquisition_settings.png! !acquisition_all_focuses.png! |