Project

General

Profile

Actions

Falcon acquire image » History » Revision 1

Revision 1/6 | Next »
Anchi Cheng, 07/22/2014 08:09 PM


Acquire images on Falcon II

Preparation in TUI/TIA

Acquire Gain/Dark References

Follow FEI's instruction in obtaining Gain/Dark References. If you plan to save frames, make sure you repeat the process with "Store intermediate images" option on.

Recommended dose rate

The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera.

Preset Recommendation

An example can be found in Pre-MSI_Set-up

Updated by Anchi Cheng over 10 years ago · 1 revisions