Anchi Cheng, 07/23/2014 03:36 PM


Acquire images on Falcon II

Preparation in TUI/TIA

Acquire Gain/Dark References

Follow FEI's instruction in obtaining Gain/Dark References. If you plan to save frames, make sure you repeat the process with "Store intermediate images" option on.

Recommended dose rate

The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera.

Preset Recommendation

An example can be found in Pre-MSI_Set-up

Updated by Anchi Cheng over 10 years ago · 2 revisions