h1. Acquire images on Falcon II
2
3
h2. Preparation in TUI/TIA
4
5
h3. Acquire Gain/Dark References
6
7 3 Anchi Cheng
Follow FEI's instruction in obtaining Gain/Dark References.
8 1 Anchi Cheng
9 3 Anchi Cheng
If you plan to save frames, make sure you repeat the process with "Store intermediate images" option on.
10
In addition, take your Gain/Dark Reference at the dose rate of your data collection when using this option.  The linearization feature in Falcon software does not work when "Store intermediate images" is on.  Take an image at a long exposure time after reference image acquisition to confirm that the references are indeed valid in this case.
11
12 1 Anchi Cheng
h2. Recommended dose rate
13
14
The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera.
15 3 Anchi Cheng
If you have to use lower dose rate, check the validity of the gain reference.
16 1 Anchi Cheng
17
h2. Preset Recommendation
18
19
An example can be found in [[Pre-MSI_Set-up#Preset-Design-Example-for-Falcon-II-on-Krios]]