Falcon acquire image » History » Version 3
Anchi Cheng, 08/08/2014 11:31 AM
1 | 1 | Anchi Cheng | h1. Acquire images on Falcon II |
---|---|---|---|
2 | |||
3 | h2. Preparation in TUI/TIA |
||
4 | |||
5 | h3. Acquire Gain/Dark References |
||
6 | |||
7 | 3 | Anchi Cheng | Follow FEI's instruction in obtaining Gain/Dark References. |
8 | 1 | Anchi Cheng | |
9 | 3 | Anchi Cheng | If you plan to save frames, make sure you repeat the process with "Store intermediate images" option on. |
10 | In addition, take your Gain/Dark Reference at the dose rate of your data collection when using this option. The linearization feature in Falcon software does not work when "Store intermediate images" is on. Take an image at a long exposure time after reference image acquisition to confirm that the references are indeed valid in this case. |
||
11 | |||
12 | 1 | Anchi Cheng | h2. Recommended dose rate |
13 | |||
14 | The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera. |
||
15 | 3 | Anchi Cheng | If you have to use lower dose rate, check the validity of the gain reference. |
16 | 1 | Anchi Cheng | |
17 | h2. Preset Recommendation |
||