Falcon acquire image » History » Version 4
Anchi Cheng, 08/08/2014 11:42 AM
1 | 1 | Anchi Cheng | h1. Acquire images on Falcon II |
---|---|---|---|
2 | |||
3 | h2. Preparation in TUI/TIA |
||
4 | |||
5 | h3. Acquire Gain/Dark References |
||
6 | |||
7 | 4 | Anchi Cheng | Follow [[Falcon_gain_correction]] |
8 | 3 | Anchi Cheng | |
9 | 1 | Anchi Cheng | h2. Recommended dose rate |
10 | |||
11 | The optimal dose rate is likely between 3 - 3.5 electrons/pixel/frame or 50-60 electron/pixel/s based on available literature of high resolution structures using this camera. |
||
12 | 3 | Anchi Cheng | If you have to use lower dose rate, check the validity of the gain reference. |
13 | 1 | Anchi Cheng | |
14 | h2. Preset Recommendation |
||
15 | |||
16 | An example can be found in [[Pre-MSI_Set-up#Preset-Design-Example-for-Falcon-II-on-Krios]] |