Faraday Cup Measurment Notes » History » Version 1
Jim Pulokas, 10/04/2010 03:16 PM
1 | 1 | Jim Pulokas | h1. Faraday Cup Measurment Notes |
---|---|---|---|
2 | |||
3 | Picoammeter setup: |
||
4 | * plug in |
||
5 | * wait 1 hour |
||
6 | * activate ZCHK (maybe already when turned on) (should say ZC on screen) |
||
7 | * plug into faraday holder |
||
8 | * set to lowest range (2nA) |
||
9 | * activate ZCOR to set zero (should say ZZ on screen) |
||
10 | * set auto range if desired |
||
11 | * deactivate ZCHK to begin measuring |