The camera modes are configured as separate cameras as follows.
- Super Resolution should be double the size of Linear and Counting.
- zplane will be relative to any other cameras in your system. Anything below K2 should have a lower zplane.
- to make the images acquired from the camera to have the standard Leginon orientation, the camera configuration in Digital Micrograph will need a rotation and flip (270 degree rotation and a horizontal flip on our camera). Therefore, the height is longer than width in the configuration below
[Gatan K2 Linear] class: dmsem.GatanK2Linear zplane: 50 width: 3710 height: 3838 [Gatan K2 Counting] class: dmsem.GatanK2Counting zplane: 50 width: 3710 height: 3838 [Gatan K2 Super] class: dmsem.GatanK2Super zplane: 50 width: 7420 height: 7676
If you have an Orius camera with K2, set its zplane below K2 like this:
[camera2] class: dmsem.GatanOrius zplane: 49