Gatan K2 Instruments cfg » History » Revision 4
Revision 3 (Anchi Cheng, 10/31/2013 05:34 PM) → Revision 4/7 (Anchi Cheng, 10/31/2013 05:38 PM)
The camera modes are configured as separate cameras as follows. * Note that Super Resolution should be double the size of Linear and Counting. * Also note that zplane will be relative to any other cameras in your system. Anything below K2 should have a lower zplane. * to make the images acquired from the camera to have the [[Leginon Image Orientation|standard Leginon orientation]], the camera configuration in Digital Micrograph will need a rotation and flip (270 degree rotation and a horizontal flip on our camera). Therefore, *the height is longer than width* in the configuration below <pre>[Gatan K2 Linear] class: dmsem.GatanK2Linear zplane: 50 width: 3710 height: 3838 [Gatan K2 Counting] class: dmsem.GatanK2Counting zplane: 50 width: 3710 height: 3838 [Gatan K2 Super] class: dmsem.GatanK2Super zplane: 50 width: 7420 height: 7676</pre> If you have an Orius camera with K2, set its zplane below K2 like this: <pre> [camera2] class: dmsem.GatanOrius zplane: 49 </pre>