The camera modes are configured as separate cameras as follows.
* Super Resolution should be double the size of Linear and Counting.
* zplane will be relative to any other cameras in your system. Anything below K2 should have a lower zplane.
* to make the images acquired from the camera to have the [[Leginon Image Orientation|standard Leginon orientation]], the camera configuration in Digital Micrograph will need a rotation and flip (270 degree rotation and a flip around the vertical axis on our camera). Therefore, *the height is longer than width* in the configuration below
<pre>[Gatan K2 Linear]
class: dmsem.GatanK2Linear
zplane: 50
width: 3710
height: 3838
[Gatan K2 Counting]
class: dmsem.GatanK2Counting
zplane: 50
width: 3710
height: 3838
[Gatan K2 Super]
class: dmsem.GatanK2Super
zplane: 50
width: 7420
height: 7676</pre>
If you have an Orius camera with K2, set its zplane below K2 like this:
<pre>
[camera2]
class: dmsem.GatanOrius
zplane: 49
width: 2048
height: 2048
</pre>