Hardware Troubles » History » Version 3

Eric Hou, 06/24/2010 11:59 AM

1 1 Amber Herold
h1. Hardware Troubles
2
3 3 Eric Hou
h2. Preset manager fails to send a preset to the microscope
4 1 Amber Herold
5 3 Eric Hou
Why: If Tecnai microscope is busy for reasons such as correcting a pole touch error, Leginon can not comunicate with the scope until the latter is free.
6 1 Amber Herold
7 3 Eric Hou
Solution: Make sure the scope is not busy any more and send the preset to scope manually to resume data collection.
8 1 Amber Herold
9 3 Eric Hou
Note: The immediate imaing condition may be wrong since manual preset sending does not include extra image shift from the target, but it should recover in later process.
10 1 Amber Herold
11
h2. Leginon can not activate GIF slit
12
13
Why: Unknown
14
15
Try these:
16
17
* Toggle the filtered button in the EFTEM menu a few times.
18
19
* Send tomography preset and then the square preset to scope.
20
21
* Make sure Gatan's GIF tracking program is off.
22
23
There is no fix for this bug.
24
25
h2. Image acquired at the wrong stage position
26
27
Commonly why: dirty or worn compustage parts
28
29 2 Amber Herold
Solution: Call for scope service
30
31 3 Eric Hou
* If this happens during grid atlas data collection, it has no impact on targets selected on it. However, there is no way to patch to the intended location easily.
32 2 Amber Herold
33 3 Eric Hou
* If this happens while a target is reacquired for drift correction, the drift correction will be wrong. There is no fix except to "declare drift" in Drift Manager to force another drift corection as soon as possible.
34 2 Amber Herold
______
35
36
[[General operation problems|< General operation problems]] | [[Troubles with Imaging - Troubleshooting|Troubles with Imaging >]]