Project

General

Profile

Hitachi install » History » Version 7

Anchi Cheng, 07/29/2021 03:35 PM

1 1 Anchi Cheng
h1. Hitachi 7800 specific installation
2
3
h2. Install myami package on the associated camera computer
4
5
* Instead of installing myami package on HHT scope-controlling PC, the access can be through the computer controlling the digital camera.
6
7
* No special supporting package needed.
8
9 7 Anchi Cheng
h2. hht.cfg
10 1 Anchi Cheng
11
hht.cfg records some instrument specific parameters that are required to convert the instrument output to SI units that Leginon depends on.
12
13
The values in pyscope/jeol.cfg.template is based on NYSBC HT7800 equipped with TVIPS XF416.  You may start with these and copy it
14 3 Anchi Cheng
to pyscope/hht.cfg or other [[locate global config directory on Windows|valid myami configuration file locations]].
15 1 Anchi Cheng
16
h2. instruments.cfg
17
18
[scope]
19
class: hitachi.HT7800
20
cs: 3e-3
21 6 Anchi Cheng
22
* Cs for HR and HC modes are 3.0 mm and 5.2 mm, respectively.  You will most likely use HR mode on the final image.
23 1 Anchi Cheng
24
* You will also be adding the camera in the same instruments.cfg since this file is located on the camera computer in this instruction.
25
26 7 Anchi Cheng
_____________
27
h2. Next Step: [[hitachi setup|Setup and calibrations]]
28 1 Anchi Cheng
h2. Testing with pyscope
29
30
In python command
31
<pre>
32
from pyscope import hitachi
33
t =hitachi.HT7800()
34
t.getHighTension()
35
</pre>
36
37
You should get the current gun high tension value in Volts.