Project

General

Profile

Pausing and Aborting during data collection » History » Version 7

Anchi Cheng, 11/28/2012 03:13 PM

1 1 Amber Herold
h1. Pausing and Aborting during data collection
2
3 4 Eric Hou
Users may need to pause leginon data collection when filling cryo-stage and cold trap or making adjustment to image or beam shift of a preset or even perform certain microscopy alignment. Pausing is possible when leginon is waiting for user input at target finder level. Alternatively, pausing can be made in any acquisition node with the pause tool.
4 1 Amber Herold
5
6 4 Eric Hou
The pause tool (as well as the abort tool), when clicked, does not interrupt acquisition until the next possible point that is at the end of an acquisition sequence for a target. When the play tool is clicked at a later point of time, leginon should move the stage and image shift to that defined by the next image. Here are some examples of the valid pausing
7
points.
8 1 Amber Herold
9 4 Eric Hou
*  At the end of an acquisition sequence for a target in the middle of a target list that may also be one of the queued list.
10
*  At the end of an acquisition sequence for a target at the end of a target list that may also be one of the queued list.
11 1 Amber Herold
*  Before a focus target is repeated after DriftManager is done monitoring drift.
12
13
Currently, you can not pause at the following points:
14 4 Eric Hou
*  Within the preset list of an acquisition target.
15
*  Before any target is processed at the node.
16 1 Amber Herold
17
h2. Pausing in MSI application
18
19 4 Eric Hou
For those using MSI-type of application, we recommand the following pausing point:
20 1 Amber Herold
21
Case 1: Exposure Targets are NOT queued
22 4 Eric Hou
*  At Hole or Subsquare (MSI-Raster) node, click on the "Pause" tool any time and it will pause before the start of next hole in the same square.
23 1 Amber Herold
24
Case 2: Exposure Targets are queued
25 3 Anchi Cheng
26 4 Eric Hou
*  In Focus node and click on the "Pause" tool while the Exposure target is being acquired. Leginon will then pause at the end of the exposure sequence before the next Focus target is processed.
27 1 Amber Herold
28 3 Anchi Cheng
h2. Abort Targets already submitted
29
30 5 Amber Herold
See http://emg.nysbc.org/boards/6/topics/146#message-187
31 6 Amber Herold
32 2 Amber Herold
______
33
34 7 Anchi Cheng
[[Do and Do Not in leginon|< Do and Do Not in leginon]] | [[Restarting procedure to continue data acquisition|Restarting procedure to continue data acquisition >]]
35 2 Amber Herold
36
______