h2. Testing feature finding algorithm
6
7
8
9
The default values in the RCT node may not work for your sample, trials should be made
10
to optimize the parameters. The good parameters should find hundreds, but not thousands of
11
features in your untilted image for good results.
12
13
14
*  Presets Manager> send the preset used in "Centered Square" to scope at the center
15
of a square.
16
17
18
*  RCT/toolbar> Click on the "acquire" button at the far right to acquire a test image
19
of which the feature finding algorithm will be applied after the acquisition.
20
21
22
*  RCT> Check the result of the numbers of feature found.  A few hundred but not
23
thousands would be a good settings.
24
25
26
*  RCT/Settings> Adjust Min Feature Size (most sensitive) and/or Max Feature Size or
27
the filters and then repeat 2 and 3 to get better results.
28
29
30
*  Target templates are applied to the valid raster points to create the final
31
acquisition and focus targets. The template can be constant, meaning that it is defined
32
relative to the image coordinate. The template can also be convolved with the raster
33
points which will create multiple targets, each defined relative to the coordinate of
34
the valid raster points.
35
36
37
38
39
*  Min/Max Feature Size: if  < 1, it means faction of the image area.  If > 1,
40
number of pixels.
41
42
43
*  List of Tilts to Collect are in the order of data collection
44
45
46
*  RCT node has its own drift monitoring after each tilt at the transformed focus
47
target position.  Therefore, drift threshold and drift preset need to be defined.
48 2 Amber Herold