Project

General

Profile

RCT node set-up » History » Version 4

Anchi Cheng, 10/12/2010 11:46 AM

1 1 Amber Herold
h1. RCT node set-up
2
3 4 Anchi Cheng
h2. Acquisition Settings that is unique to this node:
4
* Move Type: Must be stage position or modeled stage position
5
* [[Iterative Stage Movement]] Option is ON
6
* Navigator Acceptable Tolerance should be set so that sufficient overlap between tilt pairs can be achieved.
7 1 Amber Herold
8
h2. Testing feature finding algorithm
9
10
The default values in the RCT node may not work for your sample, trials should be made
11
to optimize the parameters. The good parameters should find hundreds, but not thousands of
12
features in your untilted image for good results.
13
14
15 3 Amber Herold
1.  Presets Manager> send the preset used in "Centered Square" to scope at the center
16 1 Amber Herold
of a square.
17
18
19 3 Amber Herold
2.  RCT/toolbar> Click on the "acquire" button at the far right to acquire a test image
20 1 Amber Herold
of which the feature finding algorithm will be applied after the acquisition.
21
22
23 3 Amber Herold
3.  RCT> Check the result of the numbers of feature found.  A few hundred but not
24 1 Amber Herold
thousands would be a good settings.
25
26
27 3 Amber Herold
4.  RCT/Settings> Adjust Min Feature Size (most sensitive) and/or Max Feature Size or
28 1 Amber Herold
the filters and then repeat 2 and 3 to get better results.
29
30
31 3 Amber Herold
5.  Target templates are applied to the valid raster points to create the final
32 1 Amber Herold
acquisition and focus targets. The template can be constant, meaning that it is defined
33
relative to the image coordinate. The template can also be convolved with the raster
34
points which will create multiple targets, each defined relative to the coordinate of
35
the valid raster points.
36
37
38
39
40
*  Min/Max Feature Size: if  < 1, it means faction of the image area.  If > 1,
41
number of pixels.
42
43
44
*  List of Tilts to Collect are in the order of data collection
45
46
47
*  RCT node has its own drift monitoring after each tilt at the transformed focus
48
target position.  Therefore, drift threshold and drift preset need to be defined.
49 2 Amber Herold
50
51
52
______
53
54
[[Summary of this application - MSI-RCT|< Summary of this application]] | [[Presets for MSI-RCT applications|Presets for MSI-RCT applications >]]
55
56
______