Project

General

Profile

Setup Diffraction mode TEM instrument » History » Version 3

Anchi Cheng, 12/06/2019 03:47 PM

1 1 Anchi Cheng
h1. Setup Diffraction mode TEM instrument
2
3 2 Anchi Cheng
On microscope PC, with myami-ed git branch of installation
4 1 Anchi Cheng
5 2 Anchi Cheng
h2. Edit Instruments.cfg used by leginon.  Typically in C:\Program Files\myami\
6
7 1 Anchi Cheng
Add diffraction instrument as another tem.  The class name is your normal instrument, for example, Glacios, with Diffr prefix.  Therefore, add
8
<pre>
9
[difftem]
10
class: fei.DiffrGlacios
11
cs: 2.7e-3
12 2 Anchi Cheng
</pre>
13
14
h2. Add items required for the instrument and workflow to your fei.cfg.  This should have exists with instruments.cfg
15
16 3 Anchi Cheng
Decide where you will put all the AutoIt script executable.  For example: *C:\Program Files\AutoIt3\*
17 2 Anchi Cheng
Copy the items from your pyscope/fei.cfg.template to make sure you have the right syntax.
18
19
h3. Under [camera] module, add
20
<pre>
21
AUTOIT_TUI_ACQUIRE_EXE_PATH = C:\Program Files\Autoit3\TuiAcquire.exe
22
AUTOIT_TIA_EXPORT_SERIES_EXE_PATH = C:\Program Files\Autoit3\TiaExportSeries.exe
23
AUTOIT_TIA_EXPORTED_DATA_DIR = C:\Users\supervisor\Documents\FEI\TIA\Exported Data
24
</pre>
25
26
h3. Add [beamstop] module and items needed at the end of some module.
27
<pre>
28
[beamstop]
29
AUTOIT_IN_EXE_PATH =  C:\Program Files\Autoit3\BeamstopIn.exe
30
AUTOIT_OUT_EXE_PATH = C:\Program Files\Autoit3\BeamstopOut.exe
31 1 Anchi Cheng
</pre>