Project

General

Profile

Setup and Calibration for jeolcfg » History » Version 16

Anchi Cheng, 07/14/2015 09:17 PM

1 1 Anchi Cheng
h1. Setup and Calibration for jeol.cfg
2
3 2 Anchi Cheng
All setup and calibrations are done on the computer that has access to JEOL's External COM through temext.dll.
4 1 Anchi Cheng
5
h2. Copy from pyscope/jeol.cfg.template to pyscope/jeol.cfg
6
7 3 Anchi Cheng
The current template is based on a JEM-3200FSC
8 2 Anchi Cheng
9 3 Anchi Cheng
h2. Modify [tem option] section in pyscope/jeol.cfg according to your scope
10 1 Anchi Cheng
11 3 Anchi Cheng
h3. Using PLA for image shift
12 1 Anchi Cheng
13 13 Anchi Cheng
Image shift is normally achieved with IS1.  However, PLA is a sometimes a better choice as it does not induces beam tilt at
14
the specimen, although its range is more limited.  When an omega filter is part of the scope, this also makes sure the beam not
15
affected at the entrance of the filter.
16 1 Anchi Cheng
17 13 Anchi Cheng
h3. CL3 RELAXATION MAG
18
19
The center of the beam on some scopes, such as JEM3200FSC at NYSBC, moves whenever CL3 is adjusted.  In such as case, a more stable position can be reproduced if the beam intensity is cycled back and forth through cross-over through a programable procedure called "CL3 Relaxation", found in TEMCON/Option/Lens Relaxation if available.  Leginon emulate that.
20
21
If you have this particular problem, set this option to a magnification where the procedure will be performed whenever the magnification is changed to a value larger than the "cl3 relaxation mag" value.  Typically this is set at a value below the magnification for fa,fc, and final exposure so that these become more stable.
22
For example:
23
<pre>
24
CL3_RELAXATION_MAG = 5000
25
</pre>
26
27 3 Anchi Cheng
h3. Lens Series within MAG1 and LOW MAG modes
28
29
There are distinct sets of objective and projection lens settings within MAG1 mode that gives
30
different scale and neutral values.  We assign each of these lens series (LS) numbers, starting at 1 at
31
the lowest mags. Similarly, we also named the lowest LOWMAG mode lens series LM1 and assign a
32
different deflector scale to them.
33
34
The order of magnifications is therefore roughly
35
36
<pre>
37 14 Anchi Cheng
LM1->the rest of LOWMAG->LS1->LS2->LS3->LS4->the rest of MAG1
38 3 Anchi Cheng
</pre>
39
40
* To simply the mapping of magnification to the projection modes, MAG1 will be used if a magnification
41
can be achieved by either LOWMAG and MAG1*
42
43 5 Anchi Cheng
These four configurations like these therefore define the boundary of each lens series within the mode.
44 3 Anchi Cheng
<pre>
45 9 Anchi Cheng
LS1_MAG_MAX = 4000
46 1 Anchi Cheng
LS2_MAG_MAX = 25000
47 14 Anchi Cheng
LS3_MAG_MAX = 80000
48
LS4_MAG_MAX = 100000
49 3 Anchi Cheng
LM1_MAG_MAX = 200
50
</pre>
51
52
h2. Run pyscope/scalecalibrator.py to get values for different items in jeol.cfg
53
54 8 Anchi Cheng
With the main viewing screen down do these at the computer with temext.dll registered.
55 4 Anchi Cheng
56 3 Anchi Cheng
# Open pyscope/jeol.cfg with a text editor and leave it open so that you can enter values as you go.
57
# Double click pyscope/scalecalibrator.py to start it.
58 6 Anchi Cheng
# The script will first make the scope go through all magnifications in LOWMAG and MAG1 range to determine valid magnifications.
59 10 Anchi Cheng
# Enter the magnifications at which to calibrate.  Make sure that you include at least one magnification in each lens series that you will use.
60
# Follow the instruction to shift each coil or mechanism by the amount indicated.
61
62
For example, for image shift with IS1,  It says
63
<pre>
64
Preparie to calibrate IS1:
65
Waiting for you to setup the initial condition
66
</pre>
67
The initial condition means a visible feature at an easily remembered position such as the central dot on the main screen so that a shift can be recognized.
68
The distance between the major ticks of the linear scale bar on the main viewing screen corresponds to 1 cm at the film camera (i.e., nominal magnification)
69
70
After "hit a key to continue", this original IS1 value is recorded.  You are then asked to
71
<pre>
72
Move IS1 by 1.0 cm on Main Screen in x direction
73
</pre>
74
This means using x-axis IS1 knob to move the image feature by the amount.  The major ticks of the scale bar on the main viewing screen corresponds to 1 cm at the film camera height.
75
76
The new IS1 value is saved when you hit a key to continue.
77
78
The output include
79
<pre>
80
[def]
81
IMAGESHIFT_SCALE%LS2%X=8e-8
82
</pre>
83
84
You should then modify this item in jeol.cfg according to the result.
85
86
*I do not know which values will change with lens series, nor can we assume that x and y coils have the same scale.  Please let me know what you get.*
87 1 Anchi Cheng
88 10 Anchi Cheng
*Missing calibration of unused lens series should not cause problem.  There is no need to go through all of them just for the sake of completeness.*
89 11 Anchi Cheng
90 14 Anchi Cheng
*The scale does not need great accuracy at this point, a 20% error probably is not going to cause failure since there we will refine this later within Leginon.*
91 12 Anchi Cheng
92
*The math for calculating beamtilt coil scale may not be correct.  Please report problem in later Leginon calibrations*
93
94
h2. Record standard focus and neutral deflector values
95
96
Standard focus and neutral deflector values are the references for relative movements performed in Leginon.  These values are stored in JEM scopes and will only change when engineer alignment is performed.
97
98
# Open pyscope/jeol.cfg with a text editor and leave it open so that you can enter values as you go.
99
# Double click pyscope/standard.py to start it.
100
# The script will first make the scope go through all magnifications in LOWMAG and MAG1 range to determine valid magnifications.
101
# Follow the instruction to go through each magnification that you will access.
102
# At the end the program will summarize the values which you can copy into jeol.cfg
103
104 1 Anchi Cheng
*You may notice as I have that at LS2 (and probably above) the standard defocus is unchanged at different mags.  In this case, put only the standard focus value for the lowest magnification so that the defocii can be modified together in this lens series*
105 13 Anchi Cheng
106
h2. Set up stage backlash options
107
108
Backlash correction on JEOL scopes may be different individually.  These options allows tuning of them.
109
110 1 Anchi Cheng
There are two kinds of backlash correction.  Full correction send the stage to a temporary position at a fix distance and direction relative to the final position first, before moving to the final.  The temporary position used in "reduced correction" is at half the distance of the final movement. Its purpose is to save time.
111
112
BACKLASH_SCALE:  This has separated X, Y values.  They defines the size of backlash correction (in meters) the stage makes when full correction is performed.  To determine the proper value, observe the movement of the stage when you use TEMCON program "Operation/Specimen Position" to recall a position from memory.
113 14 Anchi Cheng
114
BACKLASH_LIMIT%FULL: The smallest movement in meters at which a full correction will be performed. TEMCON program does this at all distance, but it is time consuming.  We switch to reduced method at 10 um, i.e. BACKLASH_LIMIT%FULL = 1e-5
115
116
BACKLASH_LIMIT%REDUCED: The smallest movement in meters at which a reduced correction will be performed.  We currently use reduced method down to at 0.5 um, i.e. BACKLASH_LIMIT%FULL = 5e-7
117
118 15 Anchi Cheng
h2. Enter STAGE_LIMIT and ACCURACY of the stage movement
119 14 Anchi Cheng
120 15 Anchi Cheng
axises are X,Y,Z,A, and B. XYZ are in meters, AB are in radians
121 14 Anchi Cheng
122 15 Anchi Cheng
Stage movement becomes random once it is below a particular value.  This is scope dependent.  You can specify a different values if different from the default.  Please adjust the [[Iterative Stage Movement]] Tolerance accordingly.  It may be necessary to use larger stage tilt autofocus focus sequence.  The notes in [[Adding_a_new_focus_sequence_to_a_focus_node]] is useful for designing this.
123 14 Anchi Cheng
124 15 Anchi Cheng
Jeol stage XY accuracy from our experience is in the range of 0.1-0.2 um.
125 16 Anchi Cheng
126
h2. Examples
127
128
Here is an example of jeol.cfg.  This comes from NYSBC JEM3200FSC It has energy filter and automated aperture but we do not yet align the slit during the experiment.  We also have a special lens program for 2000x which gives effective mag of 750x used for DE-20 sq mag.  Because of that it belongs to its own sub-lens series.
129
130
The file attached is jeol-3200fsc.cfg