3 | |||
4 | |||
5 | h2. Taking final exposures at very high magnification |
||
6 | |||
7 | |||
8 | |||
9 | Hysteresis of image shift at very high mags such as 200 kx can be reduced by setting the |
||
10 | following configuration |
||
11 | |||
12 | |||
13 | * Presets Manager/Settings> |
||
14 | |||
15 | |||
16 | Cycle Magnification Only = no |
||
17 | |||
18 | |||
19 | 3 | Alex Kadokura | * If [[Queuing Example 1 - Exposure Targeting|queuing is used at exposure targeting]], |
20 | 1 | Amber Herold | |
21 | |||
22 | Exposure Targeting/Settings> |
||
23 | |||
24 | |||
25 | Declare drift when queue submitted = yes |
||
26 | |||
27 | |||
28 | |||
29 | |||
30 | |||
31 | |||
32 | |||
33 | h2. Avoiding drift correction that uses LM sq |
||
34 | preset |
||
35 | |||
36 | |||
37 | |||
38 | If the grid is flat and the offset of targets is small enough that it can be corrected |
||
39 | 3 | Alex Kadokura | in the hole image, we recommand using the <[[Queuing Example 1 - Exposure Targeting|Queuing option at |
40 | Exposure Targeting]]. The sq preset is still used after Stage Z adjustment to correct |
||
41 | 1 | Amber Herold | the hole targets but will not be used when queued exposure targets are processed. |
42 | |||
43 | |||
44 | If the offset of targets is minimal even after stage Z adjustment, set the following to |
||
45 | avoid using ancestor images obtained in LM for drift management purpose, queuing or |
||
46 | not. |
||
47 | |||
48 | |||
49 | * Target Adjustment/Settings> |
||
50 | |||
51 | |||
52 | Minimum Magnification =Lowest Magnification you want to allow the node to use for |
||
53 | determining the new target. |
||
54 | |||
55 | |||
56 | |||
57 | |||
58 | |||
59 | |||
60 | |||
61 | h2. Minimize the use of image shift for the final exposure (Approach 1) |
||
62 | |||
63 | |||
64 | |||
65 | Large image shift can cause sufficient beam tilt that creates problems such as bad |
||
66 | autofocus, beam shift, and loss of resolution (but probably only if better than 4 angstrum |
||
67 | is needed). Image shift is used in Leginon for targeting exposure because it is much more |
||
68 | accurate than a single movement by the specimen goniometer, even after it is modeled for |
||
69 | mechanical periodicity. |
||
70 | |||
71 | |||
72 | The accuracy of the stage movement is lower when it moves a long distance. Therefore, |
||
73 | using iterative stage movement can improve the targeting, and therefore reduce the amount of |
||
74 | image shift required for the final exposure. To use this feature, change the following |
||
75 | preference in Hole node (or Subsquare node in MSI-Raster). |
||
76 | |||
77 | |||
78 | It is important to know that in order to check whether the precision is reached, image |
||
79 | is taken at the preset at which the target is selected on (in this case, sq preset), so the |
||
80 | dose should be kept at minimal. On our microscope 0.2 micron precision can be obtained |
||
81 | within 2 to 3 moves. |
||
82 | |||
83 | |||
84 | Hole/Settings/Image Acquisition> |
||
85 | |||
86 | |||
87 | * Mover=navigator |
||
88 | |||
89 | |||
90 | * Navigator Target Tolerance = 2 e-7 m (or whatever tolerance you like): This sets the |
||
91 | goal for multiple movement |
||
92 | |||
93 | |||
94 | * Navigator Acceptable Tolerance = 1 e-6 m (or whatever tolerance you like): If |
||
95 | further movement causes an increase rather than a decrease of targeting accuracy, this |
||
96 | value determines whether the target is aborted or not. |
||
97 | |||
98 | |||
99 | * Final Image Shift = No |
||
100 | |||
101 | |||
102 | |||
103 | |||
104 | Navigation/Settings/Error Checking and Correction> |
||
105 | |||
106 | |||
107 | * Preset cycle after each move = yes if your sq preset is in LM mode and you don't |
||
108 | mind wait a little longer; = no if you don't have hysteresis problem staying in sq |
||
109 | preset for the unknown amount of time during the iterative move. |
||
110 | |||
111 | |||
112 | |||
113 | |||
114 | |||
115 | |||
116 | |||
117 | h2. Minimize the use of image shift for the final exposure (Approach 2) |
||
118 | |||
119 | |||
120 | |||
121 | A different approach is being developed to minimize image shift while maintaining final |
||
122 | targeting accuracy-A combination of stage movement and image shift. This approach can only |
||
123 | be used in limited cases such as targets selected for the tomography node or exposure node, |
||
124 | and is only experimental in the latter case. |
||
125 | |||
126 | |||
127 | Tomography or Exposure/Settings/Image Acquisition> |
||
128 | |||
129 | |||
130 | * Move Type = modeled stage position |
||
131 | |||
132 | |||
133 | * Mover=navigator |
||
134 | |||
135 | |||
136 | * Navigator Target Tolerance = 1 e-7 m (or whatever tolerance you like) |
||
137 | |||
138 | |||
139 | * Navigator Acceptable Tolerance = 1 e-6 m (or whatever tolerance you like) |
||
140 | |||
141 | |||
142 | * Final Image Shift = Yes |
||
143 | |||
144 | |||
145 | |||
146 | |||
147 | |||
148 | |||
149 | |||
150 | h2. Optimize autofocusing sequence |
||
151 | |||
152 | |||
153 | |||
154 | The focus sequence in Focus and Z Focus nodes can and should be optimize for specific |
||
155 | cases. The activated sequence in the default setting is good for a flat holey grid and is |
||
156 | meant for high resolution imaging at 50k x or higher. If accurate defocus is not important, |
||
157 | focusing once per grid square may be sufficient, for which all focus sequence in Focus node |
||
158 | can be deactivated and the same autofocus step can be added to Z Focus node instead. |
||
159 | |||
160 | |||
161 | There are two methods to determine the defocus automatically : Stage tilt (Equivalent to |
||
162 | stage wobbling) and Beam tilt. There are also two possible ways for correcting the defocus |
||
163 | measured: Defocus (Equivalent to turning the focusing knob on the scope and reset the |
||
164 | defocus) and Stage Z (Moving the stage to the zero defocus height). |
||
165 | |||
166 | |||
167 | The following observations at NRAMM may help you determine what is the best to |
||
168 | use |
||
169 | |||
170 | |||
171 | * Both defocus determination method requires calculation correlation between two |
||
172 | images. Therefore, the magnification and the location of the autofocusing should include |
||
173 | area with contrast. |
||
174 | |||
175 | |||
176 | * The higher the magnification, the more accurate the defocus determination but the |
||
177 | smaller the range of defocus the defocus determination can handle. |
||
178 | |||
179 | |||
180 | * Beam Tilt is more reproducible and preferred method to determing defocus in HM |
||
181 | mode. |
||
182 | |||
183 | |||
184 | * Stage Tilt (Wobbling) is the preferred method to determine defocus in LM |
||
185 | mode. |
||
186 | |||
187 | |||
188 | |||
189 | |||
190 | 3 | Alex Kadokura | MSI-Raster Chapter contains an example of an [[Improving Autofocusing|alternative Z Focus node focus sequence]] for grid that show little contrast at |